I. Overview
R&S ® The ENV216 two-wire V-shaped artificial power supply network meets the requirements of CISPR 16-1-2 and EN 55016-1-2 for V-shaped artificial power supply networks. The simulated impedance is (50 μ H+5 Ω) | | 50 Ω in the frequency range of 9 kHz to 30 MHz, and meets the relevant requirements of ANSI C63.4 in the frequency range of 150 kHz to 30 MHz. A socket with grounding protection contacts can be connected to the device under test. Multiple models that comply with the specific connector systems of different countries/regions are available for selection. Thanks to its compact and lightweight design, R&S ® ENV216 is highly suitable for frequency conversion applications, especially for on-site interference voltage measurement of single-phase tested equipment.

IIENV216 Rod Artificial Power NetworkMain Features
(1) Frequency range: 9 kHz to 30 MHz
(2) 10A power processing capacity (varies by country/region)
(3) Simulated impedance (50 μ H+5 Ω) | | 50 Ω, compliant with CISPR 16-1-2
(4)VThe artificial power network complies with the requirements of CISPR, EN, VDE, ANSI, FCC Part 15, and MIL-STD-461
(5) Calibrated according to CISPR 16-1-2
IIIENV216 Rod Artificial Power NetworkFeatures and advantages
1. Hollow Inductors and Simulated Hands: Hollow Design and Simulated Hands
R&S ® The ENV216 two-wire V-shaped artificial power supply network is based on hollow inductors and equipped with analog hands, which can simulate the influence of the user's hands when measuring interference voltage.
2. High pass filter: 150kHz high pass filter can be selected through a switch
To prevent the measurement receiver from being overloaded due to high levels in the low-frequency spectrum, a 150kHz high pass filter can be inserted and used.
3. Attenuator: Built in 10dB attenuator
R&S ® The ENV216 two-wire V-shaped artificial power network is equipped with a 10 dB attenuator, which can provide standard impedance regardless of the input attenuation of the measurement receiver.
4. Pulse limiter: Built in pulse limiter (can be turned off)
The built-in pulse limiter (can be turned off) protects the measurement receiver input from high-energy RF interference pulses.
5. Remote control
TTL level remote control (compatible with Rohde&Schwarz measurement receivers)
TTL control input is driven by a controller or a Rohde&Schwarz measurement receiver and can be used for remote control of phase selection and activation of high pass filters in automatic testing systems. Optocouplers can prevent external interference.