Dynamic wax cleaning and anti wax agent evaluation deviceProduct Introduction:
This instrument is used for evaluating the anti wax effect of wax cleaning agents. By controlling the temperature difference between the paraffin oil solution and the wax deposition tube, the paraffin deposits on the wax deposition tube. The wax deposition amount of the paraffin diesel solution with and without the addition of a wax removal agent is measured on the wax deposition tube, and the wax removal rate of the wax removal agent is calculated. This instrument is designed with a dual circulation system, allowing for two sets of experiments to be conducted simultaneously with and without the addition of anti wax agents.
The instrument consists of a heating and cooling source, a wax deposition tube, a low-temperature chamber, a high-temperature chamber, a test solution tank, a stirrer, a circulation pump, a circulation tube, a flow control valve, a flow meter, a vent valve, a timer, etc.
Dynamic wax cleaning and anti wax agent evaluation deviceTechnical Indicators:
1.Heating and cooling source: a fluid that provides a certain temperature for a low-temperature chamber
2.The temperature control range is 20~50 ° C, which can be adjusted arbitrarily, and the control accuracy is ± 1 ℃
3.Wax tube: The outer diameter of the wax tube is Φ 14mm, the inner diameter is Φ 10MM, and the length is 150mm
4.Wax tube material: 45 # steel, aluminum alloy, stainless steel (1CR18NI9 π) Wax tube quantity: 10 pieces of each material
5.Low temperature chamber: The low-temperature chamber is equipped with wax tubes, which do not have cooling function and rely on heating and cooling sources to maintain a constant temperature of the fluid. Made entirely of stainless steel.
6. High temperature chamber: used to raise the temperature of paraffin diesel solution, using water bath heating method, the working temperature can be adjusted from room temperature to 0 ℃, and the temperature control accuracy is ± 1C.
7. Install the test solution tank, circulation pump, flow control valve, vent valve, circulation pipe, etc. in high temperature water to ensure that wax does not form outside the paraffin pipe. The water bath is made of all stainless steel
8. The test solution tank is installed in a high-temperature room and can hold a liquid volume of π∂ oM. It is equipped with an electric stirring device, and the stirring electric device is installed above the test solution tank. The tank body and stirring are both made of 1cR18N9T steel
9. Circulating pump: This pump is a submersible magnetic pump with a head of 22M and a flow rate of 0.1M3/H7. Flow regulation: high-precision flow regulation is adopted, and the flow rate can be adjusted arbitrarily
10. Flow meter: using glass rotor flow meter, flow range: 4~40L/H
11. The timer can be set at any time from 0 to 4H to control the working time of the circulation pump