Welcome Customer !

Membership

Help

Shandong Hongde Industrial Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Dust particle counter

NegotiableUpdate on 05/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Sampling flow rate of 28.3L/min; 5% (1.0CFM) $r $n particle size channels 0.3 μ m, 0.5 μ m, 1.0 μ m, 5.0 μ m, 10.0 μ m, 25.0 μ m $r $n laser source laser diode (with a continuous lifespan of up to 100000 hours)

Product Details

SX-L310A/HAD-L310A is a new generation of portable touch screenDust particle counterThe sampling flow rate is 28.3L/min ± 5%, which can conveniently, quickly, and effectively detect suspended particles in the purified environment. The new design weighs only 4.6kg, which is ergonomically designed for a lighter weight and suitable for one person operation. The instrument adopts a 7-inch color touch screen, with a beautiful and atmospheric appearance. The body is made of all SUS304 stainless steel material, and the surface is treated with wire drawing; Adopting a DC brushless blower, it operates continuously and stably with ultra-low noise, and is energy-saving and environmentally friendly; Built in digital flow detection sensor, dynamic and precise control of sampling flow; The power supply can be provided by AC power or by the built-in large capacity lithium battery, which is convenient for mobile detection in places without AC power. The instrument is easy to operate and has a user-friendly interface, making it suitable for rapid testing of cleanroom cleanliness in pharmaceutical and electronic factories.


model SX-L310A/HD-L310A
sampling flow 28.3L/min ±5% (1.0CFM)
Particle size channel 0.3μm,0.5μm,1.0μm,5.0μm,10.0μm,25.0μm
laser light source Laser diode (with a continuous lifespan of up to 100000 hours)
Imported fan Continuous and stable operation, energy-saving and environmentally friendly
Verification standards JJF 1190 - 2008 , GB / T 6167 - 2007
Repetitive relative error ≤10%FS
Particle size distribution error ≤±30%
Particle concentration indication error ≤±30%FS
Maximum sampling concentration 35000 pieces/liter
clean-down capability ≤ 10 minutes (counting to zero for 3 consecutive times within 10 minutes, 95% confidence level)
sampling time User self setting (1-14400 seconds)
sampling delay User self setting (1-255 seconds)
number of sampling points 1~250
Sampling frequency (location) 1~50
Sampling Room 100, room names support combinations of numbers and English
working hours Fully charged for continuous testing for 7-8 hours
Data memory capacity 5 million ordinary sampling data, 1 million UCL sampling data, and 10 million audit trail data
Counting mode Cumulative number, concentration
audit trail Record user operation events, as well as the time and date of the operation
Permission Management Instrument operation requires a power on password, and independent passwords and operation permissions are set according to the different identities of the operator, auditor, and administrator
alarm function Realize particle number alarm by setting alarm parameters
Charging adapter Input 100V-240V 1.3A, output 24V 3.75A
Battery charging time 8 hours
Cleanliness level determination criteria ISO14644-1, GMP dynamic, GMP static
temperature and humidity optional
Data communication interface RS485、 Network port, USB drive export WIFI
environment Usage environment: Temperature 0~40 ℃, relative humidity 10~70% PH
Storage environment: Temperature -30~45 ℃, relative humidity 0~90% PH
Overall dimensions Length 210.5 x Width 195 x Height 231 (mm)
material SUS304 stainless steel
power Counter (15W), charger (90W)
Weight (including battery) 4.6kg