model |
LR8102 |
LR8101 |
Connectable module |
Voltage and Temperature Module M7102 Voltage and Temperature Module M7100 Power measurement module M7103 |
|
Maximum number of modules that can be connected |
10 (M7103 up to 4) |
|
Maximum number of synchronized hosts |
10 units (requiring optical connection cables) |
Unable to synchronize with the host |
power supply |
AC adapter: AC adapter Z1016 (DC 12V ± 10% drive) External power supply: DC 10 V~30 V AC power module M1100: When using power measurement module M7103, it is necessary |
|
Temperature and humidity range for use |
-10 ℃~50 ℃, below 80% RH (no condensation) |
|
Number of LAN ports |
2 |
1 |
communication interface |
Collecting data and setting recording conditions through the Logger Utility Initial setting of IP address through Logger Utility Use communication commands for setting and recording control Manually obtain data through FTP server FTP data automatic transmission (FTP client) HTTP server functionality XCP on Ethernet (TCP) NTP client functionality
Ethernet XCP (UDP) Output measurement data through UDP |
Collecting data and setting recording conditions through the Logger Utility Initial setting of IP address through Logger Utility Use communication commands for setting and recording control Manually obtain data through FTP server FTP data automatic transmission (FTP client) HTTP server functionality XCP on Ethernet (TCP) NTP client functionality |
external media |
USB flash drive, ensure normal operation: Z4006 (16 GB) Support SD/SDHC memory cards to ensure normal operation: Z4001 (2 GB), Z4003 (8 GB) |
|
External control terminal |
Pulse/logic input (1), external sampling input (1), external input-output (4), alarm output (4), GND terminal (5), CAN interface (1) |
Pulse/logic input (1), external sampling input (1), external input-output (4), alarm output (4), GND terminal (5) |
Volume and weight |
No module: approximately 80W x 166H x 238D mm (excluding protrusions), 1.5 kg When installing one M7100 module: approximately 134W × 166H × 263D mm (excluding protrusions) When installing 10 M7100 modules: approximately 620W x 166H x 263D mm (excluding protrusions) |
|
Attachment |
Precautions for use x 1, Startup guide x 1, DVD x 1 (Startup guide, detailed user manual, Logger Utility, Logger Utility user manual, CAN unit setting software, easy to use CAN unit setting software, communication command user manual) The host does not come with an AC adapter. Please choose the separately sold (optional) AC adapter Z1016, power cord L1012, or AC power module M1100. When combined with M7103, an AC power module M1100 is required. |
|
applicableModule for measuring the voltage of stacked battery cells in battery packs and conducting charge and discharge tests on battery cell production lines
module |
Voltage/Temperature Module M7100 |
Voltage/Temperature Module M7102 |
battery pack |
Suitable for 600 V to 1500 V |
Suitable for voltages below 600 V |
Maximum rated voltage to ground |
DC 1500 V, AC 1000 V (CAT II) |
DC 600 V, AC 600 V (CAT II) |
Maximum rated voltage between modules |
DC 1500 V, AC 1000 V |
DC 600 V, AC 600 V |
maximum number of channels |
15 channels/piece |
30 channels/piece |
sampling rate |
Fastest 5 ms |
Fastest 10 ms |
More details |
Modular power measurement module suitable for photovoltaic inverter production line
module |
Power measurement module M7103 |
Maximum rated voltage to ground |
DC 1000 V, AC 1000 V (CAT III) DC 1500 V, AC 1000 V (CAT II) |
Measurement range |
Voltage: 6 V to 1500 V (8 ranges) Current: 40 mA~2000 A (6-speed range, depending on sensor) |
number of channels |
3ch/unit |
sampling rate |
Fastest 5 ms |
More details |