DC withstand voltage tester
This series of instruments can detect technical indicators such as insulation strength of the tested instruments or materials, and are widely used for voltage withstand testing of electrical equipment, appliances, transformers, power lines, electric meters, motors, power plug sockets, cables, etc. It is suitable for voltage withstand testing of household appliances such as televisions, fans, recorders, refrigerators, washing machines, microwave ovens, hair dryers, as well as electronic devices such as instruments and meters. It also provides testing methods for the implementation of national safety standards, thus demonstrating its importance and becoming an indispensable instrument and equipment for scientific research institutions, metrology and testing departments, and the electronics industry to test electronic instruments and household appliances, in accordance with the GB4706.1 standard.
This series of instruments has a beautiful appearance and is easy to operate. The output voltage is displayed by a digital panel meter, which is clear, intuitive, and accurate. The whole machine technology *, reasonable layout, high accuracy, good reliability, easy maintenance and other advantages.
2、 Main technical indicators and parameters(DC withstand voltage tester)
modelparameter |
9633 |
9634 |
9635 |
9635A |
9636 |
9637 |
9638 |
test voltage |
AC:0~5 (kV) |
AC/DC:0~5(kV) |
AC:0~5 (kV) |
AC:0~5 (kV) |
AC/DC:0~10(kV) |
AC/DC:0~10(kV) |
AC/DC:0~20(kV) |
voltage accuracy |
0.5~5(kV)≤±5% |
0.5~10(kV)≤±5% |
0.5~20(kV) ≤±5% |
||||
Leakage current range |
AC:0~20 (m A) |
AC:0~20(mA) DC:0~10(mA) |
AC:0~100(mA) |
AC:0~200(mA) |
AC:0~20(mA) DC:0~10(mA) |
AC:0~100(mA) DC:0~10(mA) |
AC:0~10(mA) DC:0~10(mA) |
Leakage current accuracy |
AC:0~20 (mA) ≤±5% |
AC:0.2~20 (mA)≤±5% DC:0.2~10 (mA)≤±5% |
AC:0~100 (mA) ≤±5% |
AC:0~200 (mA)≤±5% |
AC:0.2~20 (mA)≤±5% DC:0.2~10 (mA)≤±5% |
AC:0.2~100 (mA)≤±5% DC:0.2~10 (mA)≤±5% |
AC:0.2~10 (mA)≤±5% DC:0.2~10 (mA)≤±5% |
output waveform |
50Hz sine waveform or DC |
||||||
Transformer power |
300VA |
300VA |
500VA |
1000VA |
500VA |
1000VA |
1000VA |
time control |
Digital display 1~99 (s)±5%Manual control ∞ arbitrary preset |
||||||
display mode |
Test voltage, leakage current, and time with full digital display |
||||||
Test discrimination |
Qualified/Unqualified Unqualified Sound and Light Alarm |
||||||
Environmental Requirements |
Relative humidity: ≤ 75% RH Environmental temperature: 0 ℃~40 ℃ There is no strong electromagnetic field interference source around; No large amount of dust and corrosive gases; well-ventilated |
||||||
weight |
About 10kg |
About 12kg |
About 15kg |
22 kg |
About 15kg |
About 25kg |
About 30kg |
volume |
320×175×255 |
380×280×200 |
380×365×200 |
440×448×235 |
|||
power supply |
AC 220V ±10% 50Hz±5% |
||||||