BeburConventional five parameter automatic water quality analyzerIt is a water quality conventional five parameter monitoring instrument that can continuously detect pH, conductivity, dissolved oxygen, turbidity, and temperature in real-time online. The instrument has stable performance and is easy to install and use.
This itemConventional five parameter automatic water quality analyzerReal time display, recording, storage, downloading, and transmission of various data through the BT-6208 controller, mainly used in industries such as water supply networks, municipal secondary water supply, sewage treatment plants, food preparation, beverage processing, and swimming pool water treatment. The standard RS485 communication interface adopts the standard Modbus RTU communication protocol and can also be optionally equipped with 4-20mA output or relay output.

The BT-6208 controller developed by Bebur can connect multiple sensors of different types to achieve measurement and control of various water qualities.
Diversified communication methods
4G modem for GPRS communication, allowing customers to remotely access the controller. The optional communication options for the BT-6208 controller include 4-20mA, Modbus ASClI/RTU, Modbus TCP, and other communication methods.
More sensor configurations
The BT-6208 controller has an analog signal output function proportional to the sensor signal, and the optional PID function can control and process complex water treatment process projects.
safety
Users can set their own login password, with three levels of instrument protection and remote access security protection, allowing you to use all functions of the BT-6208 controller with confidence.
Sensor maintenance reminder
The BT-6208 controller can remind users to operate and calibrate the equipment at regular intervals according to the maintenance schedule. Users can choose to configure remote access functionality.
Type: Fluorescence method
Application: Continuous online monitoring of dissolved oxygen in water
Measurement range of dissolved oxygen meter: 0-50mg/L or 0-500% saturation
Resolution: 0.01mg/L
准确性: ±0.1mg/L(0-8mg/L<1.25%>); ±0.2mg/L(8-20mg/L); ±10%(20-50mg/L)
Working temperature: 0-50 ℃
PH range: pH2-pH10
Salinity range: 0-42pppt
Temperature compensation: automatically completed by integrated thermistor
Allowable pressure: 10 bar
Response time: T90<45s, T95<60s (25 ℃)
Zero point calibration: not required
Structural materials: PVC, silicone, polycarbonate, stainless steel, delrin,ABS
Size: Diameter approximately 38mm, length 259mm
Maintenance: Manually calibrate every 3-36 months; Replace the lumiphore every 24-48 months
Interference: Alcohol>5%, *>3%, sodium hypochlorite>3%, gaseous sulfur dioxide and gaseous chlorine
Type: LED optical sensor
Calibration: Single point calibration (no need for zero point calibration)
Accuracy:<1% of the measured value or 001NTU
Linear: r2>0.99
Resolution: 0.001NTU (0-2)
Repeatability:<0.3% NTU of measured value
Working temperature: 0-50 ℃
Typical range: 0.01-20NTU 0.01-100NTU
Work pressure: 0-10Bar
Working electrode: 316 stainless steel, sapphire
Light source: IRLED, 860nm
Response time: T90>10S
Cable length: 6 meters, extendable
Dimensions: Diameter 38mm, Length 278mm
Type: Combining reference and measurement electrodes
Calibration: Use DPD for calibration
Impedance: ≤ 130MOhm
Ph range: 0-12, 0-13, 0-14
Temperature compensation: PT100
Temperature range: -5-100 ℃
Work pressure: 0-7Bar
Measurement surface: PVC/glass
Response time: T95<3S
Dimensions: Diameter 25mm, Length 190mm
Type (1): Graphite electrode
Type (2): Circular electrode
Type (3): Stainless steel electrode (high temperature type)
Circular electrode range: 0-2000000 μ s/cm
Graphite stainless steel range: 0-5000 μ s/cm
Shell material: epoxy resin/modified polyphenylene ether/stainless steel
Working temperature: 70 ℃; 105℃; 200℃
Work pressure: 7.5bar-10Bar
Temperature compensation: including
Cable length: 6 meters, extendable
Installation methods: immersion, three-way, flow tank installation
Bebur's conventional five parameter water quality automatic analyzer can also be connected to various other sensors, such as residual chlorine, total chlorine, sludge concentration, chlorine dioxide, etc ..