- Phone
-
Address
No. 60-82, North Ring East Road, Pingjiang District, Suzhou City
Suzhou Weier Experimental Supplies Co., Ltd
No. 60-82, North Ring East Road, Pingjiang District, Suzhou City
1TC-501F(III)Cold and hot cycle analyzerWork OriginalReason:
Mainly used for automatically switching and placing test samples between high and low temperature zones in water tanks, the entire switching action is completed separately by the robotic arm. tryThe sample is first immersed in a low-temperature tank for a certain period of time and then lifted out of the liquid surface by a robotic arm. It is then automatically placed into the high-temperature tank and left there for a certain period of time before returning from the high-temperature tankThe sink is put back into the low-temperature sink, and this process is counted as one cycle per unit. The placement time and circulation of the sample in two different temperature zones of the water tankThe number of times and the speed of movement of the robotic arm can be freely set through a microcomputer controller.
IITC-501F(III)Cold and hot cycle analyzerTechnical parameters:
1Temperature control range of low-temperature tank:0~100℃Constant temperature wave degree:≦±0.5℃Temperature control range of the constant temperature bath:RT~100℃Constant temperature wave degree:≦±0.5℃The accuracy is:±0.1℃The uniformity is:±0.5℃
2Low temperature tank cooling power:368WHeating power:1500W×2(Double slot)
3Effective volume of liquid tank:≥10LNet size of studio:290×190×200mm
4Adjustable range of slot to slot conversion speed:30~150rpm
5Adjustable range of dwell time in the slot:1~300S
6Adjustable range of cycle times:1~9999time
7Controller surface:LCDLCD display, capable of controlling temperature, conversion speed, dwell time, and number of cycles (upgraded)
8Conversion of support frame load:500g (can also be customized)
9External dimensions:680×460×840mm
10Power supply voltage:220V,50Hz
11Sample basket size: diameter70mm, height70mmThe circular hanging basket is made of304Stainless steel (can also be customized according to requirements)
12Equipment material: The shell is made of1.0Made of high-quality cold plate electrostatic spraying, the inner liner adopts1.0of premium304Stainless steel fabrication
IIITC-501F (III) Thermal Cycling InstrumentQuotation for conducting experiments on behalf of others:
According to the commonly used testing requirements of users, the dwell time for each slot is 25-30 seconds, the conversion time is 5 seconds, and the quotation based on the number of cycles is as follows:
1000 times: 400 yuan (about one day)
2000 times: 800 yuan (about two days)
3000 times: 1200 yuan (about three days)
5000 times: 1900 yuan (about five days)
10000 times: 3600 yuan (about 10 days)
......
The frequency can be set according to user requirements, and specific prices are welcome to be discussed by phone
Note: The commissioned experiments provided by our company are suitable for users who do not use the equipment frequently, which can greatly save them costs and avoid the waste of instruments that are often left unused.
4TC-501F (III) Cold and Hot Cycling InstrumentProduct related instructions:
This product has obtained metrological certification from professional testing institutions and holds other relevant certificates. Please feel free to contact us for inquiries.

5、 Precautions for operation and use:
1. When the device is in operation, please do not open the box door unless necessary, otherwise it may cause the following adverse consequences. If high-temperature hot air rushes out of the box during high-temperature testing, it is very dangerous.
2. If during low-temperature testing, the freezing system is prone to frosting, which affects the testing effect; If the inside of the box door still maintains high temperature, causing injury, the high-temperature air may also trigger a fire alarm, resulting in misoperation.
3. Please note that it must be safely grounded to avoid static electricity induction.
4. Avoid shutting down and restarting the refrigeration unit within three minutes.
5. When using water cooling in winter, please use tap water or treated soft water to keep the cooling system circuit smooth.
6. If a heating sample is placed inside the cold and hot cycle impact test chamber, please use an external power supply for the sample power control, and do not directly use the local power supply.