- Phone
-
Address
No. 16 Tengda Road, Shangfeng Industrial Park, Tangshan Street, Jiangning District, Nanjing City
Nanjing Huanke Experimental Equipment Co., Ltd
No. 16 Tengda Road, Shangfeng Industrial Park, Tangshan Street, Jiangning District, Nanjing City
Condensation test chambermodel |
NLX-010A |
Working size cm |
100×100×100 |
Dimensions cm |
180×130×210 |
Studio temperature range |
10℃~80℃ |
Temperature Uniformity |
≤ 2 ℃ (when unloaded) |
temperature fluctuation |
± 0.5 ℃ (when unloaded) |
temperature deviation |
≤±2℃ |
Test humidity |
100% |
Temperature range of sink |
10℃~80℃ |
time control |
0~9999H59M59S |
Box material |
The inner box is made of high-quality 304 stainless steel plate, and the outer shell is made of galvanized steel plate sprayed with plastic |
internal structure |
The bottom of the inner tank is a water tank with adjustable water temperature, ensuring that the airflow of the air conditioning cabinet can flow through the water surface; The equipment is equipped with a plastic cover that matches the size of the test chamber, ensuring that the distance between the plastic cover and the laboratory wall is 10% of the width/depth of the laboratory; There are ventilation holes on the plastic cover, which are closed during the heating stage and opened during the cooling stage. There are lead holes on the side of plastic cover 2; |
control system |
7-inch true color LCD touch screen for controlling studio temperature and humidity, sink temperature, and testing time |
protective device |
Multiple protection functions such as phase loss protection, leakage protection, and over temperature protection |
power supply voltage |
AC380V/50HZ |
Power (KW) |
12 |
Execute to meet standards |
Condensation test of K-15 together with component group in VW80000 standard DIN EN ISO 6270-2 Climate test method for condensate water DIN 50017 Condensation Water Test Climate Test GS95011-4-2004 Condensation Test |
thisCondensation test chamberIt can be used to assess and determine the adaptability of electrical, electronic products or materials to storage and use under environmental conditions where temperature cycling changes and condensation occurs on the product surface. The equipment adopts forced air circulation to maintain the uniformity of temperature in the studio. To limit the impact of radiation, the temperature difference between each part of the inner wall of the equipment and the temperature specified in the test shall not exceed 8%, and the test sample shall not be directly radiated by the heating and cooling elements inside the equipment.