Concentration gas analyzerProduct Introduction
MIX series updatesConcentration gas analyzer, capable of diluting standard gases in a certain proportion to output a set concentration gas, can be connected
Accept four types of standard gases. Calibration instruments used for gas sensors, gas analyzers, and other products. This product is based on gas passing through MFC (Mass Flow Control)
The principle of manufacturing equipment is based on ISO 6145/7 and other standards. Equipped with a capacitive touch screen LCD interface, all functions are easy to manage
Use it. Each mass flow controller is calibrated by a traceable primary standard factory. Each mass flow controller is equipped with traceable standards
Factory calibration with NIST (National Institute of Standards and Technology) traceable calibration certificate.
Concentration gas analyzerProduct features
High precision quality flow controller, with an accuracy better than 1%
Efficient mixing device and excellent anti adsorption pipeline design, suitable for dilution and gas distribution of easily adsorbed and reactive compounds
Wide dynamic range (conventional dilution ratio 1:1~1:100000);
scalable
Fast gas flow control (stabilization time<100 ms)
Preheating time ≤ 5 minutes
Flexible configuration. Standard gas can switch up to 4 gases in a single channel
It has storage function and can store 30 calibration points. Having a sequence
Calibration function, capable of storing 30 calibration sequences
Equipped with timing function, the instrument automatically starts sequence calibration
Equipped with Ethernet, RS485, RS232 interfaces, capable of remote control of instrument operation
The instrument has an automatic cleaning function, which can automatically clean the pipeline
Concentration gas analyzerProduct Parameters
Flow accuracy: 0.8% set point (optional 0.6% set point)
Flow repeatability: 0.2% full scale
Dilution ratio: 1:1~1:1000 (customizable)
Response time:<=600ms
Preheating time:<=5 minutes
Dilution gas flow range: 0~10SLPM (other ranges can be selected)
Standard airflow range: 0~100SCCM (other ranges can be selected)
Pipeline material: 316L stainless steel/PFA/PTFE
Power supply: 200~240V, 50/60Hz
Power supply:<=50W
Communication interface: 1 * Ethernet; 1*RS485; 1*USB; 2*RS232
Communication protocol: MODBUS-RTU
Working temperature range: 0~50 ℃