Chloride ion online monitoring instrument/Chloride ion concentration detectorCL-3000Measurement principle:
The ion electrode measures the potential of a specific ion in a solution, which is measured with reference to a reference electrode with a constant built-in potential. The difference in activity of the ion in the solution will form a certain potential difference, which is related to the ion concentration. Thus, analyze and measure the ion.
Chloride ion online monitoring equipment/Chloride ion concentration detectorCL-3000Technical parameters:
Measuring range:1.8~35500ppm(pH2~12Between);
Temp:0.0~80.0℃
part distinguish Rate:0.01ppm;Temp:0.1℃
Instrument accuracy: ±2%Measurement value ±2B. C.D.
Temp:±0.5%F.S±1Digit
temperature compensation-10.0~80.0℃ automatic compensation
Operating Temperature -10℃~40℃
display screen Double row three and a half digit numberLCDdisplay
analog output A set of isolated4~20mAoutput Factory settings: corresponding0.0~100.0ppm
relay output two groupsRelay ON/OFF
inch size 96×96×90mm; Opening size1/4DIN; weight0.96Kg
Industrial composite electrode:0-80℃, withstand voltage0.6Mpa;10KWarm compensation, standard wire length 10 Mi,3/4thread
Regarding electrode installation:
Chloride ion online monitoring instrument/Chloride ion concentration detectorCL-3000Correction operation steps:
Connect the machine to the electrode correctly; Prepare two concentrations of ion standard solution(10.0ppmand100ppm);
1Low concentration correction
Wash the electrode thoroughly with deionized water and use filter paper to absorb the moisture. Place the electrode in10.0ppmIn the standard solution and Stir thoroughly. After the measured values to be displayed have stabilized, zero point correction can be performed. The steps are as follows:
① Press Key screen Appearing on the left side of the upper rowSTANDReminder, accompanied byBUFCompared to standard liquid types(1.00 ppm /10.0 ppm /100ppmor.100%)Alternating flashing[At this time, you can press or ↓ Key to change standard liquid type];
② Press Confirm the type of standard solution with the key and enter the calibration program.
③ When the upper rowSTANDWhen flashing, it indicates that the instrument is undergoing automatic calibration;
④ WhenSTANDWhen the measurement interface disappears and appears, zero point calibration is completed.
2High concentration correction:
Used for electrodes Deionize thoroughly and wash thoroughly, then use filter paper to absorb dry water;Put in high concentration100ppmIonic standard solution and Stir thoroughly. Slope correction can only be performed after the displayed measurement values have stabilized. The steps are as follows:
① Press key 2 On the secondary screen Appearing on the left side of the upper rowSLOPETips;
② Press The key,SLOPEWhen flashing,Indicates that the instrument is undergoing automatic calibration;
③ WhenSLOPEWhen the measurement interface disappears and appears, the slope correction is completed.
【 Electrode Storage Ionic selective electrodes can be stored in low concentration standard solutions during measurement gaps. If not used for a long time, it must be cleaned thoroughly, covered with a protective cover, protected from direct sunlight, and stored dry.
【 Working environment The electrode needs to work in a flowing liquid (otherwise it will affect the electrode life).
Attachment: [Preparation Method of Electrode Standard Solution]
1. Transfer1mLof1000ppmSpecification of ion concentration standard solution and99mLDeionized water, stir evenly, and add an appropriate amountISA, sample solution andISAratio 50:1, can be formulated into10.0ppmStandard solution.
2. Will90mlAdd deionized water10mLThe specifications are1000ppmMix the ion concentration standard solution evenly and add an appropriate amountISA, sample solution andISAratio 50:1, can be formulated into100.0ppmStandard solution.