- Phone
-
Address
No. 10, Lane 255, Xiaotang Road, Fengxian District, Shanghai
Shanghai Chuanna Experimental Instrument Co., Ltd
No. 10, Lane 255, Xiaotang Road, Fengxian District, Shanghai
Carbon dioxide shaker CN-CO2-300A suspension cell culture technology parameters:
model |
CN-CO2-300A |
CN-CO2-300B |
CN-CO2-300C |
number of layers |
single layer |
double-layer |
three floors |
Rotation frequency/accuracy |
10~300rpm /±1rpm |
||
Oscillation amplitude |
Ф26mm |
||
Reachable capacity |
Single layer approximately 250ml * 25 or 500ml * 16 or 1000ml * 9 or 2000ml * 5 |
||
Standard Configuration |
Universal fixture (optional fixed fixture or other fixtures) |
||
Shake board size |
Approximately 520 * 420mm |
||
timing range |
0-999.9 hours |
||
Temperature control range/accuracy |
4℃~60℃/±0.1℃ |
||
Carbon dioxide probe |
Imported infrared carbon dioxide sensor |
||
Range of carbon dioxide concentration |
0~20% |
||
Humidification method |
The natural humidification of the water pan is ≥ 90% (RH%), and this parameter is displayed as uncontrolled |
||
power supply |
220V 50HZ |
||
Carbon dioxide shaker operation manual
1、 Preparation before startup
Check the appearance of the equipment, confirm that the power and gas connections are normal, without looseness or damage.
Open the valve of the carbon dioxide cylinder and adjust the pressure reducing valve to 0.1-0.2 MPa to ensure that there is no leakage in the gas circuit.
Add distilled water or deionized water to the sink until the water level reaches the mark to avoid dry burning.
2、 Operation steps
Connect the power, press the power button, and enter the main interface.
Set parameters: Adjust the temperature (room temperature -50 ℃), speed (0-300rpm), CO ₂ concentration (0-20%), and running time through the touch screen, confirm and click "Start".
After the parameters stabilize, place the culture bottle symmetrically on the shaker tray and close the hatch.
3、 Shutdown and maintenance
After running, click "Stop" to turn off the device power and gas cylinder valve.
Regularly replace the water in the sink, clean the trays and cabin walls, and keep the equipment dry.
Check the air circuit sealing and motor operation status every quarter, and replace aging components in a timely manner.
Attention: Avoid frequent opening and closing of the cabin door during operation to prevent parameter fluctuations; If there is an abnormal alarm, stop the machine immediately for troubleshooting.
Main features:
1. A combination of one, two, or three layers stacked together to provide users with the maximum usable space structure with the smallest footprint, resulting in a smaller footprint and larger usable space;
2Carbon dioxide shaker CN-CO2-300A suspension cell cultureEach layer is independently controlled and can operate simultaneously at different temperatures and speeds, or run one, two, or three layers as needed;
3. LCD touch screen, set temperature, speed, time, and measured temperature, speed, and remaining time are displayed on the same interface without switching interfaces, making observation more direct
Observation; The encryption and locking function of the operation interface prevents duplicate operations and human errors;
4. Having data recording function, it can record data from the past three months; There is a USB interface that can export and save the above data;
5. After the oscillation experiment is completed, it automatically switches to use in the incubator, runs quietly, and is designed without fluorine;
6. Functions such as over temperature alarm, over speed alarm, scheduled shutdown, power recovery, parameter encryption, temperature correction, speed correction, etc.
7. Equipped with power-off recovery function, the device can automatically resume operation according to the original set program after the external power supply suddenly loses power and then resumes power
8. Constant temperature control system, fast response, high temperature control accuracy, air duct ventilation, soft working room wind speed, uniform temperature;
9. Closed type compressor, fluorine free environmentally friendly refrigerant, low noise, good refrigeration effect, ensuring long-term stable operation of equipment in low temperature state;
10. Hollow tempered glass door, convenient for observing the inside of the box from various angles without opening the door at any time;
11. Adopting imported infrared CO2 sensors, automatic control of CO2 concentration, sensitive response, precise and stable concentration;
12. Equipped with servo motors, precise speed control, good high-speed performance, and strong stability;
13. The box is equipped with ultraviolet sterilization lamps that can regularly disinfect the box with ultraviolet radiation, thereby preventing contamination during cell culture;
14. The studio adopts water tray natural evaporation humidification, and the humidity is directly displayed by the instrument