Carbon dioxide flame gas thermometer optris CTlaser F2

NegotiableUpdate on 01/03
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The optris CTlaser F2 $r $noptris CTlaser F2 infrared laser thermometer provides a special measurement wavelength for the temperature measurement of CO2 flame gas. This makes this thermometer suitable for processes such as combustion, waste incineration, or chemical reactors. The temperature range of the $r $n infrared thermometer is 200 ° C; C to 1650 C, And stainless steel probes can withstand 85 without cooling; C. Optional cooling and protection equipment ensures accurate temperature measurement even in harsh environments.

Product Details

Carbon dioxide flame gas thermometer optris CTlaser F2

The optris CTlaser F2 infrared laser thermometer provides a special measurement wavelength for temperature measurement of CO2 flame gas. This makes this thermometer suitable for processes such as combustion, waste incineration, or chemical reactors.

The temperature range of the infrared thermometer is 200 ° C to 1650 ° C, while the stainless steel probe can withstand 85 ° C without cooling. Optional cooling and protection equipment can ensure accurate temperature measurement even in harsh environments.

638470390888262093804.jpg

Product Specifications

measurement parameters


Temperature range (measured through setting keys or software)

200°C ~ 1,450°C (CTlaser F2)
400°C ~ 1,650°C (CTlaser F2H)

spectral range

4.24 µm

Optical resolution (90% energy)

45:1

System accuracy (ambient temperature=23 ± 5 ° C)

± 1% (Note 1) (Note 2)

Repetitive accuracy (ambient temperature=23 ± 5 ° C)

± 0.5% or ± 0.5 ° C (Note 2)

Temperature resolution (digital)

0.1K

Exposure time (90% signal) (Note 3)

10 ms

Emissivity/gain (adjusted by setting keys or software)

0.100 ~ 1.100

Transmittance/gain (adjusted by setting keys or software)

0.100 ~ 1.100

Signal processing (specific parameters can be adjusted through setting keys or software)

Peak hold, valley hold, average, extended hold threshold, and lag function

Basic Parameters


environmental level

IP65(NEMA-4)

ambient temperature

Probe: -20 ° C~85 ° C (laser on at 50 ° C); Electronic box: 0 ° C~85 ° C

Storage temperature

Probe: -40 ° C~85 ° C; Electronic box: -40 ° C~85 ° C

relative humidity

10~95%, no condensation

Vibration (sensor)

-IEC 60068-2-6 (sinusoidal waveform)/- IEC 60068-2-64 (wideband noise)

Impact (sensor)

-IEC 60068-2-27 (25G and 50G)

weight

Probe: 600 g; Electronic box: 420 g

Electrical parameters


analog output

0/4-20mA、0-5/10V、 Thermocouple J, K

alarm output

24 V/50 mA (normally open)

optional

Relay: 2 x 60 V DC/42 V AC; 0.4 A; optical isolation

Digital output (optional)

USB, RS232, RS485, Modbus RTU (serial communication protocol), Profibus DP (fieldbus multifunctional slave design), Ethernet

Output impedance

MA Max 500 Ω (DC 5-36 V); MV Min 100 k Ω (pure impedance load); Thermocouple 20 Ω

signal input

Programmable function input, used for external emissivity adjustment/environmental temperature compensation, triggering (holding function reset)

cable length

3m (standard), 8m, 15m

power supply

DC 8-36 V

Current consumption

Max 160 mA

Laser 635 nm

1mW, Open/close through electronic box or software

Note 1

The temperature of the object is higher than 300 ° C

Note 2

ε (emissivity)=1, response time: 1s

Note 3

Dynamic adaptation at low signal levels

638467150992770907938.jpg

Carbon dioxide flame gas thermometer optris CTlaser F2

Application areas of optris CTlaser F2 infrared thermometer

The Optris CTlaser F2 infrared thermometer is designed specifically for measuring the temperature of combustion gases. This thermometer is capable of measuring carbon dioxide gas up to 1650 ° C during combustion processes. This device is suitable for different processes such as waste incineration or chemical reactors.

model
temperature range
temperature measurement accuracy
response time
distance coefficient
spectral response
OPTCTLF2L
200~1450℃
±1%
10ms
45:1
4.24μm
Optctlf 2 h
400~1650℃
±1%
10ms
45:1
4.24μm
OPTCTLF6L
200~1450℃ ±1%
10ms 45:1
4.64μm
Optctlf 6 h
400~1650℃
±1%
10ms
45:1
4.64μm
Operating Temperature
Probe: -20~85 ℃, electronic box: 0~85 ℃
analog output
0/4-20mA, 0-5/10V, J/K type thermocouple signal output
digital output
USB、RS232,RS485,CAN-Bus,Profibus DP, Ethernet 可选
emissivity
0.100~1.100 software settings
Connecting cable length
3m, 8m, 15m optional
signal processing
Peak holding, valley holding, average value.
power supply
8~36VDC
application
Temperature of flue gas containing CO2 or CO gas in heating furnaces, boilers, and chemical reaction furnaces





CTlaser F2-4.jpg