Faster, brighter, and more features
CSU-X1 laser confocal scanning unitIt is a high-speed model of the CSU series and a tool for live cell imaging.
Scanning speed up to 2000 fps
Micro lens enhanced Nipkow disk scanning
Filter wheel with six filter positions (high-end model)
Replaceable dichroic mirror block and emission filter
3 configurations: single camera model, dual camera model, and clear field of view model
CO2Reduce emissions by 40%
Principle of Micro Lens Enhanced Nipkow Rotary Scanning Technology

CSU-X1 laser confocal scanning unitA Nipkow disk containing approximately 20000 pinholes and a disk containing the same number of microlenses (used to focus the excitation laser into each corresponding pinhole) are mechanically fixed with a motor, and when rotated, the field of view is scanned by a fast grating with approximately 1000 laser beams. The pinhole and micro lens patterns are designed exclusively by Yokogawa Electric to optimize grating scanning. Using CSU-X1 for multi beam scanning can not only improve scanning speed, but also significantly reduce photobleaching and phototoxicity, as multiple excitations only require lower levels of laser power on the sample to excite fluorescence.
Detailed introduction
faster
Scanning speed up to 2000fps
Yokogawa Electric's proprietary filter wheel with six filter positions moves to adjacent positions at a relatively fast speed of 33 milliseconds.
brighter
By using the newly developed beam shaping lens, the excitation power efficiency can be doubled*1Allowing the use of lower power lasers and reducing camera exposure time.

By reducing background noise by one-third, the signal-to-noise ratio can be increased by three times*1Expand the ability of true low light imaging. By using highly efficient dichroic mirrors and emission filters, the image is significantly brighter*1.

More features - Dual camera option * 1
Users can use two cameras to simultaneously image two different emission ranges, or selectively use one of the two cameras installed that is more suitable for the current experimental requirements. For each camera port, users can choose to install a high-speed filter wheel (optional). In addition to the standard C-mount adapter, adapters for 8X8 EMCCD cameras and F-mount cameras are also available (optional).

More Features - Clear View Path Options * 2
It allows users to use a camera for confocal imaging through CSU-X1 and bright field (non confocal) imaging through a bypass optical path.
More features - replaceable mirror block and replaceable filter
Easy to replace dichroic mirror block and emission filter.

*1. It is necessary to use the commercially available dichroic mirror selected by the user for synchronous multi-color imaging.
*Due to spatial interference, the bright field option is not suitable for certain microscope settings. Please consult its applicability in advance.
High end model (6-digit filter wheel)

High end model (12 position filter wheel))

Basic model

Host*1 |
Host+ |
Host+ |
|
Confocal scanning method |
Micro lens enhanced Nipkow disk scanning |
||
scanning speed |
Selection: 1500-5000rpm (standard) |
||
External Synchronization |
Synchronize scanning speed through pulse signals |
||
excitation wavelength |
405~647nm |
||
Another port |
- |
bright field |
Dual cameras |
dichroic mirror |
Optional*4 |
||
Mirror switching |
Automatic 3-channel |
||
Laser beam input |
AFC connector (polished 8 degrees) |
||
optical fiber |
Standard single-mode polarization maintaining fiber of Yokogawa Electric Corporation*5 |
||
Filter wheel |
Filter wheel with six filter positions |
||
Emission filter |
Optional*4 |
||
Control Panel |
Open/close laser shutter |
||
external control |
Through the RS-232C interface of the control unit |
||
Microscope bracket |
C mount adapter |
||
Operating environment |
15-40 ℃/20-75% relative humidity |
||
energy consumption |
24VDC maximum 1A |
||
energy consumption |
Input: 100-240VAC ± 10%, 50 or 60Hz ± 3Hz, maximum 75W |
||
External dimensions*6 |
175(W)×328.5(H)×301.5(L)mm |
175(W)×328.5(H)×301.5(L)mm |
175(W)×328.5(H)×301.5(L)mm |
Weight*7 |
8.9kg |
11.7kg |
13.0kg |
*1. Provide a control unit (for the filter wheel) and a filter wheel.
*Provide two control units (one filter wheel and one bright field) and one filter wheel.
*3 options available
*4 does not include filters (excitation filter, emission filter, dichroic mirror). If necessary, please consult in advance.
*5 are provided with each CSU-X1 host.
*6 does not include protruding parts. Including the filter wheel.
Host*1 |
Host+ |
Host+ |
|
Confocal scanning method |
Micro lens enhanced Nipkow disk scanning |
||
scanning speed |
Selection: 1500-5000rpm (standard) |
||
External Synchronization |
Synchronize scanning speed through pulse signals |
||
excitation wavelength |
405~647nm |
||
Another port |
- |
bright field |
Dual cameras |
dichroic mirror |
Optional*4 |
||
Mirror switching |
Automatic 3-channel |
||
Laser beam input |
AFC connector (polished 8 degrees) |
||
optical fiber |
Standard single-mode polarization maintaining fiber of Yokogawa Electric Corporation*5 |
||
Filter wheel |
Filter wheel with 12 filter positions |
||
Emission filter |
Optional*4 |
||
Control Panel |
Open/close laser shutter |
||
external control |
Through the RS-232C interface of the control unit |
||
Microscope bracket |
C-mount adapter |
||
Operating environment |
15-40 ℃/20-75% relative humidity |
||
energy consumption |
24VDC maximum 1A |
||
energy consumption |
Input: 100-240VAC ± 10%, 50 or 60Hz ± 3Hz, maximum 75W |
||
External dimensions*6 |
258(W)×329.8(H)×213.4(L)mm |
259(W)×374.3(H)×248(L)mm |
309.8(W)×329.8(H)×392(L)mm |
Weight*7 |
7.8kg |
10.6kg |
12.2kg |
*1. Provide a control unit (for the filter wheel) and a filter wheel.
*Provide two control units (one filter wheel and one bright field) and one filter wheel.
*3 options available
*4 does not include filters (excitation filter, emission filter, dichroic mirror). If necessary, please consult in advance.
*5 are provided with each CSU-X1 host.
*6 does not include protruding parts. Including the filter wheel.
Host*1 |
Host+ |
Host+ |
|
Confocal scanning method |
Micro lens enhanced Nipkow disk scanning |
||
scanning speed |
Selection: 1500-5000rpm (standard) |
||
External Synchronization |
Optional*4 |
||
excitation wavelength |
405~647nm |
||
Another port |
- |
bright field |
Dual cameras |
dichroic mirror |
Optional*5 |
||
Mirror switching |
Manual Channel 1 |
||
Laser beam input |
AFC connector (polished 8 degrees) |
||
optical fiber |
Standard single-mode polarization maintaining fiber of Yokogawa Electric Corporation*6 |
||
Filter wheel |
- |
||
Emission filter |
Optional*5 |
||
Control Panel |
Open/close laser shutter |
||
external control |
-*4 |
||
Microscope bracket |
C mount adapter |
||
Operating environment |
15-40 ℃/20-75% relative humidity |
||
energy consumption |
24VDC maximum 1A |
||
energy consumption |
Input: 100-240 VAC ± 10%, 50 or 60Hz, up to 75W |
||
External dimensions*6 |
175(W)×328.5(H)×213.4(L) mm |
259(W)×373(H)×213.4(L)mm |
308.5(W)×328.5(H)×213.4(L)mm |
Weight*7 |
7.5kg |
10.0kg |
10.0kg |
*1. Provide a control unit (for the filter wheel) and a filter wheel.
*Provide two control units (one filter wheel and one bright field) and one filter wheel.
*3 options available
*4. The control unit is required for speed control and external synchronization.
*5 does not include filters (excitation filter, emission filter, dichroic mirror). If necessary, please consult in advance.
*6 is provided with each CSU-X1 host.
*7 does not include protruding parts.
Control unit/filter wheel
control unit | |||
type |
Used for 6-digit filter wheel (F1) |
Used for 12 position filter wheel (F2) |
Used for bright vision (B1) |
Operating environment |
15-40 ℃ 20-75% relative humidity |
||
energy consumption |
Input: 100-240VAC ± 10%, 50 or 60Hz, maximum 200VA |
||
Overall dimensions |
213(W)×132(H)×438(L) mm |
||
Weight (kg) |
5.2 kg |
5.2 kg |
5.1 kg |
6-position filter wheel for CSU-X1 | |
Operating environment |
15-40 ℃ 20-75% relative humidity |
energy consumption |
- |
Overall dimensions |
112(W)×226(D)×100(H) |
weight |
1.9 kg |
12 position filter wheel for CSU-X1 | ||
Operating environment |
15-40 ℃ 20-75% relative humidity |
|
energy consumption |
- |
|
Overall dimensions |
154(W)×98(D)×154(H) |
|
weight |
2.4 kg |
|
Comparison between target product and reference product
Target product: CSU-X1 |
Reference product: CSU22+filter wheel |
|||||||
Energy (MJ) |
CO2Emissions (kg) |
NOx emissions (g) |
SOx emissions (g) |
Energy (MJ) |
CO2Emissions (kg) |
NOx emissions (g) |
SOx emissions (g) |
|
Raw materials |
2,335.5 |
132.5 |
337.1 |
130.1 |
2,610.0 |
148.0 |
374.9 |
147.1 |
component |
10,510.2 |
494.5 |
1,423.4 |
452.8 |
15,329.2 |
728.1 |
2,061.7 |
671.7 |
Processing and assembly |
1,542.9 |
68.7 |
214.1 |
56.8 |
2,259.8 |
100.7 |
313.6 |
83.1 |
logistics |
2,717.7 |
185.5 |
210.8 |
471.1 |
3,211.7 |
218.9 |
252.3 |
553.9 |
consume |
36,713.5 |
1,635.9 |
5,094.8 |
1,351.5 |
66,087.0 |
2,944.8 |
9,171.0 |
1,126.4 |
process |
-569.9 |
-31.9 |
-75.7 |
-33.6 |
-612.8 |
-34.5 |
-81.2 |
-37.4 |
Total |
53,249.8 |
2,485.3 |
7,204.4 |
2,428.8 |
88,884.8 |
4,106.0 |
12,092.2 |
3,851.5 |
Functional factors |
40.1% |
39.5% |
40.4% |
36.9% |
||||
Note: The value of reference products is calculated based on functional factors. Functional factor: 2.16