E-mail
service@hrtech.cn
- Phone
-
Address
No. 409 Xinfeng Road, Xukou Town, Wuzhong District, Suzhou City, Jiangsu Province
Suzhou Hongrui Purification Technology Co., Ltd
service@hrtech.cn
No. 409 Xinfeng Road, Xukou Town, Wuzhong District, Suzhou City, Jiangsu Province
CLJ-B110-Pro Laser Dust Particle CounterMainly used to measure the size and number of dust particles per unit volume of air in clean environments, it can simultaneously monitor the number and changes of particles in 8 particle size channels,The minimum particle size range can reach 0.1 μ m.This product adopts an 8-inch high-resolution color touch screen, which is easy to operate and canQuickly reflect the status of the instrument,The shell made of 316 stainless steel material can withstand various disinfectants.
CLJ-B110-Pro Laser Dust Particle CounterMeet the requirements of Chinese CGMP regulations, and all collected data comply with the requirements of PDA21CFR Part 11 regulations.Can be widely applied toElectronic, optical, chemical, food, cosmetics, medical and health, biological products, aerospace and other departments.
Appearance diagram:

built-in printer |
Product features and performance:



ØHigh resolution 8-inch color touch screen display;ØThe data storage capacity can reach at least 1 million sets;
ØSorting and organizing data according to customer needs increases the convenience of management;ØBuilt in printer, data can also be directly printed;
ØAccept USB data download, data cannot be edited or modified, in accordance with the FDA21CFRParl1 standard (electronic signature function);
ØReal time display of sampling room, sampling period, sampling frequency, etc;ØIt can display eight particle size channels simultaneously and detect up to 0.1 μ m;
ØThere are three display modes to choose from: table, line chart, and bar chart, making data changes more intuitive;
ØIndividual room names can be named in both Chinese and English; ØBuilt in multiple certification modes, with the option to choose from Chinese standards such as the new version of GMP and ISO14644-1;
ØEnter the cleanroom area to automatically calculate the cleanroom requirements Number of monitoring points;ØHaving user management functions, it can query, add, edit, and delete users;
ØUsers can be grouped and managed, and support the management and setting of permissions at or above the third level;ØMultiple users can be set for each level of permission;
Main parameters:
model |
CLJ-B110-Pro |
sampling flow |
28.3L/min |
Sampling channel |
0.1,0.15,0.2,0.25,0.3,0.5,1.0,5.0(μm) |
Particle size resolution |
<15% @ 0.5 μ m (compliant with ISO21501-4) |
laser light source |
laser diode(semiconductor laser source),, lifespan>35000 hours |
Upper limit of sampling concentration |
1000000 pieces/cubic foot (35000 pieces/L) |
Data memory capacity |
Can store at least 1 million sets of sample data; Acceptable USB data download, data cannot be edited or changed Complies with the FDA21CFRParl1 rule standard
|
contact method |
MODBUS TCP |
Data communication interface |
USB (USB drive export and upper computer software export), RJ45 socket |
sampling period |
User self setting (1 ∽ 999999 seconds) |
sampling delay |
User self setting (0-65535 seconds) |
sampling interval |
User self setting (0-65535 seconds) |
working hours |
Can work continuously for 8 hours |
Battery life |
5-6 hours |
Battery charging time |
6 hours |
power supply |
110-240V,50-60Hz, Built in rechargeable lithium-ion battery |
language |
Default Simplified Chinese version, optional English version (can be extended to other languages) |
Cleanliness level determination criteria |
GMP-2010、ISO-1999、ISO-2015 |
Room preset functions |
Countless room location numbers, points, sampling times, sampling periods, intervals, delays, alarms, etc. can be presetSet saved parameters without the need for repeated settings
|
sampling mode |
Time sampling and volume sampling modes can be set, and sampling time and time interval can be set |
Printing method |
Built in thermal printer |
working environment |
Temperature: 10-35 ℃; Relative humidity 20-75% RH; atmospheric pressure: 86-106kPa |
storage environment |
-30 ℃~ 45 ℃ Relative humidity 0-90% RH |
power |
≤50W |
Shell material |
316L stainless steel |
Dimensions (width x height x depth) |
235 × 210 × 235 (mm) (excluding air nozzle, handle, etc.) |
Weight (including battery) |
7.8Kg |
Standard Accessories |
Waiting for power sampling head, sampling head hose, self-cleaning filter, sampling bracket, thermal printing paper, Portable instrument box, lithium battery charger, data USB flash drive, user manual, original calibration report, certificate of conformity, etc |