- Phone
-
Address
Room 307, Building 4, Huixin Yuan, No. 9 Huixin West Street, Chaoyang District, Beijing
Beijing Oriental Peak Technology Co., Ltd
Room 307, Building 4, Huixin Yuan, No. 9 Huixin West Street, Chaoyang District, Beijing
High temperature heat flux sensor, high temperature radiation heat flux meter
Model: GH/HT-50
Beijing High Temperature Heat Flow SensorKey Features:
·DaqPRO 5300 data logger from Fourier Israel
Composed of the HT-50 high-temperature heat flux sensor from ITI Corporation in the United States, it is the most versatile
Can colleagues connect 4 HT-50 heat flow sensors
Beijing High Temperature Heat Flow Sensor
·Hot flow range up to 3.14 MW/m2
·The highest system resolution is 141.3 W/m2
·Working temperature up to 980 ℃ (no need for water cooling), built-in K-type thermoelectric
I am used for temperature measurement
·Maximum response time of 0.1 seconds and accuracy of 5%
·512K data storage, USB interface, rechargeable battery power supply
·*Settings, data collection and management, data analysis software,
Can achieve fast data download, online real-time retrieval, and display of data
(Numbers, bar charts, curve charts, analog instrument pointers, and digital meters)
Various forms such as tables, graphical analysis, merged data processing, and direct
Output data to EXCEL, adjust and calibrate DaqPRO settings, define new
The sensors and engineering units have numerous practical and * functions
Key Features:
·DaqPRO 5300 data logger from Fourier Israel
Composed of the HT-50 high-temperature heat flux sensor from ITI Corporation in the United States, it is the most versatile
Can colleagues connect 4 HT-50 heat flow sensors
·Hot flow range up to 3.14 MW/m2
·The highest system resolution is 141.3 W/m2
·Working temperature up to 980 ℃ (no need for water cooling), built-in K-type thermoelectric
I am used for temperature measurement
·Maximum response time of 0.1 seconds and accuracy of 5%
·512K data storage, USB interface, rechargeable battery power supply
·*Settings, data collection and management, data analysis software,
Can achieve fast data download, online real-time retrieval, and display of data
(Numbers, bar charts, curve charts, analog instrument pointers, and digital meters)
Various forms such as tables, graphical analysis, merged data processing, and direct
Output data to EXCEL, adjust and calibrate DaqPRO settings, define new
The sensors and engineering units have numerous practical and * functions
Main technical indicators:
Measurement object: Determine the total heat flux (conduction, conduction) of the high-temperature object being measured
Flow and radiation
Heat flow range/resolution:
The recorder uses a 50mV range and a 10V range
Range 2.355 MW/m2 3.14 MW/m2
Nominal resolution 141.3 W/m2 9.42 kW/m2
Heat flow measurement accuracy: 5%
Heat flow response time: 0.1 seconds
Sensor operating temperature: -250~980 ℃ (no need for water cooling)
Sensor size and cable: Φ 25.4 × 3.5mm, 1-meter long fiberglass insulated wire
Temperature measurement: K-type thermocouple, range: -250~1200 ℃, precision
Degree: ± 0.5% of full scale, resolution: 0.1 ℃
Data storage: 512KB
Sampling rate: When measuring only one heat flux: 1 time/hour to 4000 times/second
When measuring heat flux and temperature simultaneously: 1 time/hour to 1 time/second
Interface: USB 1.1
Power supply: Built in rechargeable battery, with a battery life of 40 after being fully charged
Hour, or external DC input
Standard configuration:
DaqPRO 5300 Recorder (including USB cable,
DaqLab software, charger, portable case, Chinese and English operators
Volume), HT-50 High Temperature Heat Flow Sensor 1 piece, Magnetic Installation Kit