- Phone
-
Address
334 Zijing Qixing Plaza, No. 8 Wenhua East Road, Changping District, Beijing
Beijing Yuguan Century Technology Co., Ltd
334 Zijing Qixing Plaza, No. 8 Wenhua East Road, Changping District, Beijing
Effective emissivity, 0.995
High precision 0.25%+1 ℃
Excellent stability, 1 hour+/-0.5 ° C
Rapid heating, from room temperature to 1000 ℃ in 1 hour
Large range, 50-1700 ℃
Can provide third-party verification certificates
RS 232 (standard) or RS 485 serial communication
BBC1700 medium temperature blackbody furnaceIt is a domestically developed medium temperature blackbody furnace that can be set at any temperature within the range of 50 ° C to 1700 ° C The temperature of the blackbody cavity is measured by precision thermocouples, ensuring high precision and repeatability.
The end of the cavity is closed, and the outlet diameter is 25 mm (1 "), heated by a specially designed element The temperature controller adopts the industrial standard PID algorithm to control the radiation temperature within+/-0.5 ° C
BBC1700Medium temperature blackbody furnaceAdopting resistance heating, it has a long service life and is equipped with an over temperature power cut-off system, which increases the safety performance of the system. The internal fan maintains a safe and comfortable temperature on the surface of the cabinet.
Infrared temperature sensor
infrared thermal imager
spectrometer
Sun simulator
Optical pyrometer
| test parameters | |
|---|---|
| Temperature range: | 50 . .. 1700 °C |
| temperature accuracy1: | ± 0.25% ± 1 ° C of reading |
| Temperature resolution: | 0.1 °C |
| stability2: | ± 0.5 ° C every hour |
| Uniformity: | ± 1 ° C, 1/3 diameter area from the center |
| Heating chamber shape: | Closed tube 41 mm ID X 250 mm, approximately 125 mm heating length (can be replaced on site) |
| Radiation cavity diameter: | 25 mm (1”) |
| Emissivity: | 0.995 @ 0.65 . .. 1.8 μm |
| Standard calibration method: | radiation method |
| Temperature sensor: | B-type thermocouple |
| Heating time: | From ambient temperature to 50 ° C, approximately 30 minutes 50 . 1000 ° C for approximately 80 minutes 1000 . 1500 ° C for approximately 90 minutes |
| Communication/Interface | ||
|---|---|---|
| Remote temperature setting point: | via serial port | |
| Control method: | Digital self-tuning PID controller | |
| environmental parameters | ||
| Working environment temperature: | 0 . .. 44 °C | |
| Cooling: | Fan cooling, with air inlet on the rear panel | |
| Working humidity: | Maximum, 90% RH, non condensing | |
| Dimensions (H x W x D): | 690 mm x 505 mm x 550 mm | |
| Weight: | 80 kg | |
| Electrical parameters | ||
| Power requirements: | 208 . .. 240 V AC, 50Hz, Maximum power of 3000 watts | |