Automatic cleaning viscosity meter

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The NCY-C1 automatic cleaning unit, as a subsequent configuration equipment of the NCY automatic viscosity meter, is used for the discharge of waste liquid after the automatic testing work of the automatic viscosity meter is completed, as well as for the automatic cleaning required after several hours of operation in the viscosity tube. The $r $n $r $nNCY-C1 automatic cleaning unit is compatible with all of our company's NCP automatic viscosity meters.

Product Details

NCY-C1 Automatic cleaning unit, as NCY Subsequent configuration equipment for automatic viscosity analyzer, In automatic viscosity The waste liquid discharge after the instrument completes the automatic testing work, as well as the automatic cleaning required after several hours of operation in the viscosity tube Used for washing.

NCY-C1 Automatic cleaning unit and all of our company's NCY Automatic adhesiveThe measuring instrument product is compatible.

1. waste liquid discharge

At the end of each test, It is necessary to test the solution (solvent) inside the viscosity tube Discharge, To proceed to the next step Periodic testing work, Using the waste disposal function of this unit, There is no need to remove the viscosity tube again, Just press the button The waste disposal button of the unit, It can safely discharge the waste liquid.

2. automatic cleaning

Use the viscosity tube for a period of time (such as one week)), The accumulation of residual night on the capillary wall will affect the test results When necessary, the viscosity tube needs to be cleaned, After pressing the cleaning button, The cleaning unit automatically executes the following program:

a The washing solution is automatically sent into the viscosity tube, Filled with capillaries and other places;

b Suction soaking (takes several hours));

c Discharge waste washing solution;

d Drying program.

Due to the entire process after startup, No longer requires operator intervention, Therefore, Operators and these hazardous chemicals So it was isolated.

1.Structure Introduction

C1 The cleaning unit is controlled by a controller Viscosity tube adapter component, Composed of attachments, The control unit contains a piping system Related to software, hardware, and electronic control systems.

Each controller controls the cleaning of two sets of viscosity tubes.

The control unit contains an automatic cleaning pipeline system, (Abbreviated as pipeline system, ) And software, hardware, and electronic control systems unify, (Abbreviated as electronic control system).

Pipeline system, electrical control system, and software system, All are proprietary intellectual property rights.

1.1.pipeline system

In the pipeline system, a batch of pumps and valves are installed to provide directional flow and power for liquids and gases.

1.2.electric control system

According to the experimental requirements, a series of instructions are required to automatically complete the experiment, including one for automatic cleaning Series of instructions.

1.3Viscosity tube adapter component

The appearance of the viscosity tube adapter component is shown in the figure2

Contains one subject, two rootsF4 Tube, thin tube is used for receiving and discharging waste liquid, and thick tube is used for introducing cleaning solution.

2. system connection

The entire cleaning system consists of multiple units and components, and before starting work, these units and components need to be sorted out Connect them together.

3. operation

3.1preparation

3.1.1Place it horizontally and securely on the workbench.

3.1.3Preparation work for five reagent bottles:

Fill the water bottle with purified water and add it to the top of the bottle2cmBelow is the panel when there is a shortage of water 53#lamptwinkle.

Add detergent to the detergent bottle(Chromic acid wash solution or other),Add to the bottle mouth2cmBelow, when lacking detergent, noodles board 34#lamptwinkle.Add drying aid to the drying bottle(Like acetone),Add to the bottle mouth2cmBelow, when lacking drying solution, the panel33# lampflicker

Both the waste liquid collection bottle and the wastewater collection bottle are empty bottles.

3.2cleaning operation

3.2.1Preparation before cleaning

Check the washing solution and drying solution(Like acetone)Whether the purified water is incomplete, if unsure, it is recommended to follow3.1.3want Please fill it up; Check if there is enough storage space in the waste liquid bottle and waste water bottle. If unsure, it is recommended to discharge in advance Empty.

3.2.2Insert the ventilation pipe joint onto the viscosity tube, as shown in the figure1011.

3.2.3Start cleaning

Press the cleaning button on the left unit “X1”Key, at the same timeX1The key indicator light is on, entering the cleaning program. On the right side, there is a single light Press the cleaning button “X2” Key, at the same timeX2The key indicator light is on, entering the cleaning program. After completing the cleaning steps, it will End of movement.

3.2.4Cleaning completed

After cleaning, press againX1or X2 Key, the corresponding key indicator light goes out, it is necessary to exit the cleaning program If so, drain the waste liquid bottle and waste water bottle completely.

3.2.5Interrupt cleaning

In the startup cleaning state, press the cleaning button to interrupt the unit, and the cleaning button will pop up, indicating the cleaning process Interrupt, if there is residual liquid in the viscosity tube, press the corresponding unit drain button to drain it.

3.3Drainage operation

3.3.1Start draining

Press the drain button on the left unit “P1” Key, at the same timeP1 The key indicator light is on, indicating the need for drainage and the drainage time4min. Press the drain button on the right unit “P2”Key, at the same timeP2 The key indicator light is on, indicating the need for drainage and the drainage time4min.

3.3.2Interrupt drainage

After starting the drainage, if you want to interrupt the drainage, the corresponding drainage button can pop up, and the indicator light of this button will turn off at the same time Afterwards, the drainage will stop.

4Normal testing link

4.1 During normal testing, rubber joints of different sizes are connected(As shown in the figure12)

4.2Before testing, use a dedicated sand core funnel to add the sample into the viscosity tube

Working voltage:220V(±10%),50Hz;

Overall dimensions: 200×300×950(L×W×H,mm)placed inNCYBehind the automatic viscosity analyzer;

heavy Quantity:10Kg

5Regarding the compatibility of products retained in the market

Considering that we1994Starting from the year, it took four years to develop1998Entering the market annuallyNCY Automatic viscosity For over 20 years, Yi has achieved a market share in the domestic market80%Above, in order to facilitate the technology of the aforementioned users Renovation, saving money, so our currentNCY-C Fully automatic cleaning unit, compatible with the originalNCYAutomatic adhesive The viscometer enables users of automatic viscosity meters, including both early and potential ones, to use itFor convenient useNCY-C

Fully automatic cleaning unit.