- Phone
-
Address
Jinnan Industrial Park, Jinhu County, Huai'an City, Jiangsu Province
Jinhu Jinma Automation Instrument Factory
Jinnan Industrial Park, Jinhu County, Huai'an City, Jiangsu Province
Large screen dot matrix LCD display, Chinese menu operation, simultaneous display of multiple parameters: PH/ORP value
Temperature, output current, alarm points, etc. are displayed simultaneously, which is intuitive and easy to read.
Automatic temperature compensation function.
Communication function: Equipped with RS-485 communication interface (partially compatible with MODBUS protocol) (optional),
Capable of converting to RS-232. The PH/ORP value corresponding to the 4-20 mA current output can be set.
technical indicators
1. Measurement range: -2.00~16.00 (PH), -1999~+1999mV (ORP), -10~130.0 ℃
2. Resolution: 0.01PH, 1mV, 0.1 ℃
3. 精 度: ±0.02 PH,±1mV,±0.3℃
4. Stability: ≤ 0.01pH/24h
5. PH standard solution: 6.86/4.00/9.18 7.00/4.00/10.01
6.Temperature compensation: 0-99.9 ℃ (pH)
7. PH correction range: zero point ± 1.45PH; slope ± 30%
8. Control interface: Two sets of ON/OFF relay contacts, divided into high point and low point alarm control.
9. Signal isolation output: 4-20mA isolation protection output
10. Relay: Relay hysteresis can be set arbitrarily, relay load 3A 220VAC
11. Working conditions: ambient temperature of 0-60 ℃, relative humidity ≤ 90%
12. Input impedance: ≥ 1 × 1012Ω
13. Output load: Load < 750 Ω (4-20mA)
14. Working voltage: 220VAC10%, 50/60Hz or order DC24V and DC12V power supply
15. Size: 96 × 96 × 115 mm 16. Hole size: 91 × 91mm 17. Weight: 0.9Kg