-Product Overview:
The Amtax Inter2C ammonia nitrogen monitor complies with the national standard (HJ536-2009) detection method and has the advantages of fast response speed, user-friendly operation, and wide range of measurement.
-Working principle:
Salicylic acid indophenol blue method (compliant with DIN 38406 E5 and HJ536-2009 standards). Under the action of a catalyst, NH4+reacts with hypochlorite ions and salicylate ions in an alkaline medium with a pH of 12.6 to form indophenol compounds, which appear green. Within the measurement range of the instrument, the degree of color change is directly proportional to the concentration of NH4+in the sample. Therefore, by measuring the degree of color change, the concentration of NH4+in the sample can be calculated.
-Application industry:
Used for online monitoring of ammonia nitrogen concentration in the inflow and outflow of sewage treatment plants.
-Instrument features:
Can measure NH4+ion concentration in water using a dual beam and dual filter photometer. By measuring the reference beam, the instrument eliminates the interference of factors such as turbidity and power supply fluctuations in the sample on the measurement results.
The measured values can be displayed in graphical or numerical form.
● It has functions such as automatic calibration and automatic cleaning.
Built in refrigerator to ensure the storage temperature of reagents.
● Data storage function and graphic display function.
● Use CYQ type water sample preprocessor for sample pretreatment.
● Support Chinese operation interface, better display interface for historical data.
Language: Chinese, English.
| TECHNICAL INDEX | |
|---|---|
| Measurement range: | Low range: 0.10~20.00mg/L NH4-N |
| High range: 3.00~80.00mg/L NH4-N | |
| Accuracy: | Low range of 0.10~20.00mg/L: ± (4.0%+0.10mg/L) |
| High range 3.00-80.00mg/L: | |
| When<10.00mg/L: ± (4.0%+0.20mg/L) | |
| When ≥ 10.00mg/L: ± (4.0%+0.60mg/L) | |
| Measurement interval: | 5-minute, 10min, 30min, 1h, 2h and external triggering |
| Repeatability: | ≤3% |
| Stability: | ±10% |
| Zero drift: | ±5% F.S |
| Instrument calibration: | Automatic or manual calibration |
| Automatic calibration cycle: | Closed, 0.5 days, 1 day, 2 days... 7 days optional |
| Operating temperature: | 5-40℃ |
| Sample temperature: | 5-40℃ |
| Sample pressure: | 1-5psig |
| Output: | 1-channel digital communication output |
| 2-channel 0/4-20mA analog current output | |
| Relay control: | 2-channel 24V/2A relay single pole double throw control |
| Digital output: | MODBUS or PROFIBUS |
| Power supply: | 220VAC±10%, AC50/60Hz |
| Protection level: | IP54, For indoor installation only |
| Dimensions (height x width x depth): | 1235mm x 550mm x 400mm (including refrigerator) |
| Weight: | 43kg (excluding reagents) |
| ordering guide | |
|---|---|
| Amtax Inter2C Host | |
| AT | 0.10-20.00mg/L, Single channel, no bus interface |
| AT | 0.10-20.00mg/L, Single channel, Modbus |
| AT | 0.10-20.00mg/L, Single channel, Profibus |
| AT | 3.00-80.00mg/L, Single channel, no bus interface |
| AT | 3.00-80.00mg/L, Single channel, Modbus |
| AT | 3.00-80.00mg/L, Single channel, Profibus |
| reagent | |
| LCW802 | Chemical reagents (A, B and their respective additives) |
| LCW804 | Zero point standard solution |
| LCW803 | Standard solution (5mg/L), low range |
| LCW808 | Standard solution (35mg/L), high range |
| LCW819 | Cleaning solution (2 × 250mL) |
| CYQ type water sample preprocessor | |
| CYQ-106A | 50-1000mL/min (adjustable) |