E-mail
eric@sdicelc.com
-
Phone
18911571363
-
Address
15th Floor, Building 6, Courtyard 2, Wuliqiao Second Street, Chaoyang District, Beijing
Beijing Zhongqing Guotou Environmental Protection Technology Co., Ltd
eric@sdicelc.com
18911571363
15th Floor, Building 6, Courtyard 2, Wuliqiao Second Street, Chaoyang District, Beijing
-Product Overview:
The Amtax Compact II ammonia nitrogen analyzer/detector is the second-generation online ammonia nitrogen analyzer/detector launched by Hash Company for the Chinese market based on years of research and development, production, and maintenance experience.
The Amtax Compact II ammonia nitrogen analyzer/detector can provide customers with more accurate measurement results and a more convenient operation and maintenance experience. Compliant with the national standard (HJ536-2009) for ammonia nitrogen detection method, it has the advantages of fast response speed, user-friendly operation, and wide range of measurement.
-Working principle of Amtax CompactII ammonia nitrogen online analyzer:
The Amtax Compact II ammonia nitrogen analyzer mixes the analyzed sample with the reaction reagent and measures the NH in the solution4+Conversion of ions into ammonia gas (NH)3)Ammonia gas is released from the analyzed sample. Then transfer the ammonia gas to the measuring cell containing the indicator and dissolve it again in the indicator. These reactions will cause a change in the color of the solution. The ammonia nitrogen analyzer uses colorimetric methods to measure and finally calculates the concentration value of ammonia nitrogen.
-Application industry:
The Amtax Compact II ammonia nitrogen analyzer is used for online rapid detection and analysis of ammonia nitrogen in industrial and municipal wastewater fields.
-Instrument features:
● More friendly
The Amtax Compact II ammonia nitrogen analyzer/detector supports a Chinese operating interface and a better historical data display interface, making it easier for users to operate more intuitively.
● More stable
Added a constant temperature module to ensure that the ammonia nitrogen analyzer/detector can maintain stability even in environments with temperature changes, without the need for additional constant temperature devices.
● More convenient
In the measurement gap of actual water samples, standard solution measurement can be inserted to ensure the stability of ammonia nitrogen determination/detection instrument is monitored.
● More range options
In addition to maintaining the previous generation's large range of 0.2-1200mg/L NH4+-N. The Amtax Compact II ammonia nitrogen analyzer/detector has also expanded its low range to 0.20-30.00mg/L, meeting the measurement requirements of a wide range of wastewater treatment plant outlets.
● Easy to maintain
The Amtax Compact II ammonia nitrogen analyzer/detector uses a mechanical device consisting of a peristaltic pump and a pinch valve to transport reagents and water samples. Because it does not come into direct contact with reagents, the ammonia nitrogen detector does not require frequent cleaning and maintenance, greatly reducing the waste of manpower caused by frequent component replacement.
● Automatic leak detection
The ammonia nitrogen detector is equipped with a humidity sensor, which can sense the liquid at the bottom of the ammonia nitrogen analyzer casing. If liquid is sensed, the sensor will turn off the ammonia nitrogen analyzer and display "leakage fault" in the lower left corner of the display screen. Automated leak detection can assist operators in detecting leaks early.
● Self cleaning function
The ammonia nitrogen detector has an automatic cleaning cycle that uses a solution to be expelled. The frequency of the cleaning cycle can be set through the "Settings" menu. Can ensure more accurate measurements every time and reduce manual maintenance.
-Installation of Amtax CompactII Ammonia Nitrogen Online Analyzer Instrument:
The Amtax Compact II ammonia nitrogen analyzer is designed for use in moderate dust levels, indoors, and at room temperature (10-40 ° C). The environmental standard for the instrument casing is IP54 (with the instrument door locked). Please do not install the instrument in direct sunlight and avoid water dripping onto the instrument.
The ammonia nitrogen analyzer is designed with 3 # 10 screws and fixing anchors, which can be installed on walls. If feasible, please place the instrument as close as possible to the sampling point so that the instrument can fully update the water sample in the sampling pipeline for each measurement.
During installation, please pay attention to leaving a space of no less than 20cm around the instrument for easy cleaning and pipeline connection.
-Performance parameters of Amtax CompactII ammonia nitrogen online analyzer:
● Instrument requirements
Language: Chinese, English
Shell: IP54; For indoor installation only
Dimensions (height x width x depth): 640 x 350 x 220mm
Weight: 15kg (excluding reagents)
Measurement method: Expulsion colorimetric method
● Sample requirements
Pressure entering the instrument: 1-5 PSI
Sample flow rate: 100mL/hr
Sample temperature range: 10-40 ℃
Sample entry: 0.8mm inner diameter transparent hose
Sample discharge: 1/4 inch tube
● Requirements for reagent dosage
Indicator: 1L/month
Expulsion solution and standard solution: 250mL/month
● Power supply
Power supply: 100-230 V ± 10%, AC 50/60Hz
Installation category: II
Power connection: Three hole socket connection, the socket must be reliably grounded (PE)
Safety level: Class I, protective grounding
Relay contact point: load 24V, 2A (note: do not connect to 220VAC AC mains)
Output: 2-channel analog current output: 0/4-20mA (load not exceeding 500 ohms)
Two relays: load 24V, 2A can control external circuits or issue alarm signals.
RS485 interface, MODBUS protocol
-Service products:
To ensure that customers can use the Amtax Compact II ammonia nitrogen analyzer/detector better, more confidently, and conveniently, Hash provides the following instrument service products tailored for customers:
Routine maintenance of the instrument (order number: SCAC01): safety grounding inspection, safety inspection of the power supply unit; Communication settings check; Analog current output inspection and calibration; Software version check and upgrade; Diaphragm pump testing; Pinch valve opening voltage test; Pipeline inspection; Reagent inspection; Instrument self calibration point setting; Instrument calibration; Manufacturer's testing report. One service is limited to one customer's on-site Amtax Compact II ammonia nitrogen analyzer/detector.
Instrument deep maintenance (order number: SCAC02): Routine maintenance of the instrument, pipeline maintenance, and a set of annual maintenance components (LZV149) are included as a gift. One service is limited to one customer's on-site Amtax Compact II ammonia nitrogen analyzer/detector.
Worry free guarantee plan (order number: SCAC03): Routine instrument maintenance, pipeline maintenance, data comparison (instrument standard solution inspection, water sample and Hash laboratory method comparison inspection), with a set of annual maintenance components (LZV149) and a pinch valve (LZV183) included as a gift. One service is limited to one customer's on-site Amtax Compact II ammonia nitrogen analyzer/detector.
| TECHNICAL INDEX | ||
|---|---|---|
| Measurement range: | 0.2-1200mg/L NH4-N | |
| According to the different reagents, it can be divided into the following sections: | 0.20-12.00mg/L H4-N | |
| 0.20-30.00mg/L H4-N | ||
| 2.0-120.0mg/L H4-N | ||
| 20-1200mg/L H4-N | ||
| Accuracy: | (0.20-12.00) mg/L : | (5%+0.1mg/L) |
| (0.20-30.00) mg/L: | ≤12.0mg/L:± ( 2.5%+0.1mg/L) | |
| >12.0mg/L : ±(5%+0.1mg/L) | ||
| (2.0-120)mg/L: | ±(2.5%+1mg/L) | |
| (20-1200) mg/L: | ±(5%+5mg/L) | |
| Measurement lower limit: | 0.2mg/L | |
| Measurement interval: | 13, 30 minutes, 1, 2, 4, 6, 24 hours or user-defined | |
| Signal output: | 2-channel analog current output: 0/4-20mA (maximum load not exceeding 500 ohms) | |
| Two relays: maximum load 24V, 2A can control external circuits or issue alarm signals. | ||
| RS485 interface, MODBUS protocol | ||
| ordering guide | |
|---|---|
| Host ordering | |
| Standard configuration includes: host, installation tools, one-year maintenance parts, testing report, operation manual (reagents need to be ordered separately) | |
| AC0150001 | Amtax Compact II Ammonia Nitrogen Analyzer, 0.2-12 mg/L |
| AC0150004 | Amtax Compact II Ammonia Nitrogen Analyzer, 0.2-30 mg/L |
| AC0150002 | Amtax Compact II Ammonia Nitrogen Analyzer, 2-120 mg/L |
| AC0150003 | Amtax Compact II Ammonia Nitrogen Analyzer, 20-1200 mg/L |
| Reagent ordering | |
| 2512210 | Amtax Compact II reagent kit, 0.2-12.0 mg/L |
| 2512212 | Amtax Compact II reagent kit, 0.2-30.0 mg/L |
| 2512214 | Amtax Compact II reagent kit, 2-120 mg/L |
| 2512216 | Amtax Compact II reagent kit, 20-1200 mg/L |
| Spare parts ordering | |
| LZX355 | Wall mounted installation accessories |
| LZX153 | Please refer to page 47 for ordering details of installing the Filtrax sampling and filtering system attachment |