This 500 model perfectly combines advanced technology and sturdy instruments, helping users make full use of existing manpower. The current trend is to streamline institutions, which means instrument technicians need to do more in a shorter amount of time. Therefore, instruments must be more reliable
This 500 model perfectly combines advanced technology and sturdy instruments, helping users make full use of existing manpower. The current trend is to streamline institutions, which means instrument technicians need to do more in a shorter amount of time. Therefore, instruments must be more reliable and easy to use. Emerson's Rosemount Analytical gas chromatograph is your best choice; Our equipment is highly reliable, has stable long-term performance, is easy to use, and has received widespread praise.
Environmental simulation room testing
Emerson Process Management provides the world's most comprehensive process gas chromatography testing. Each 500 type product must be operated for at least 24 hours in an environment testing room that is accessible to humans, between 0-130 ° F. Because we promise that all provided process gas chromatographs can provide reliable measurements on site.
Unparalleled measurement performance
Maximum specified accuracy (± 0.25 BTU/1000, used for environmental C6+analysis)
Wide dynamic measurement range, with component concentrations ranging from percentage to trace levels
Reliable operation within a wide ambient temperature range of -18 to 55 ° C (0-131 ° F)
Reduce operating and maintenance costs
No need for protective cover or instrument air
Low consumption of helium and electricity
Gas chromatograph injection valve and chromatographic column provide the longest warranty period
easy to use
Ethernet and Modbus connection
MON2000 software settings and diagnostics
All models have identical settings and operations - no additional training required
Power Supply:
115 VAC ±15%,220 VAC ±15%,50/60 HZ
Operating power of 100 watts, starting power of 350 watts
Environmental temperature: -18 to 55 ° C (0-130 ° F)
Shell protection level: NEMA 4
Size (excluding sampling system): 164cm high x 51cm wide x 53.5cm deep (64.5 inches high x 20 inches wide x 21 inches deep)
Installation method: no bracket installation (standard); Wall installation (optional)
Approximate weight (excluding sampling system): approximately 36 kilograms (79 pounds)
Regional security authentication options (depending on hardware):
Standard: Designed to comply with UL Class l, Zone 1, Groups C and D (Group B requires air purge option)
Optional CSA NRTL/C certification: Level l, Zone 1, Groups C and D T3B
Optional ATEX certification: EEx d llB T4amb=60 ° C, CE marking in accordance with ATEX directive (94/9/EC)
Thermostatic furnace: Non air bath furnace, maximum 85 ° C (185 ° F)
Injection valves: 6-way and 10 way diaphragm chromatography injection valves may use other types of valves, such as liquid injection plunger valves or rotary valves, depending on the application
Carrier gas: depends on the application. Usually zero order helium, nitrogen, or hydrogen at 90 PSI
Detector: Thermal Conductivity Detector (TCD); Up to two, depending on the application. Provide flame photometric detector (FPD) (please refer to FPD module sample)
Spectral peak gating options: fixed time, automatic slope sensing peak gating
Flow path: up to 12 flow paths (including calibrated flow paths)
Chromatograph control electronic unit: installed together with the chromatograph in an explosion-proof enclosure or remotely mounted on a 19 "rack
Analog inputs: 4 inputs filtered by transient protection (please note that these 4 inputs are used by the second TCD)
Analog output: Standard configuration includes 2 (up to 10 optional) 4-20 mA outputs that are not isolated
Serial communication ports: The standard configuration includes 3 serial ports, with a total of 8 optional. Provide RS-232, RS-422, RS-485 selection and Modbus protocol based on port conditions
Digital switch input: 1 gas chromatograph alarm and 4 user configurable inputs, optically isolated, with transient protection function
Digital switch output: 5 digital switch outputs can be used for alarm, optical isolation, and have transient protection function
Parallel printer port (optional): 1 parallel port for printing reports
Internal modem (optional): can be configured on-site; 300-19.2k baud rate
Transient protection has been tested and certified by C.E. and meets the highest level of IEC 801 STD (3 and 4) in Europe.