1Characteristics and main functions of the instrument
1Power on self-test,different levelsFault alarm and prompt, automatic positioning of sample tray;
2The sample area, injection valve, and sample transfer tube are all individually heated and temperature controlled;
3Through time programming, automatic functions such as pressurization, sampling, injection, analysis, and post analysis blowback cleaning can be achieved;
4. Silanizationchemical inertnesstransformProcessing, sample flow pathandallAdopting constant temperature controlAfter each injection is completed, the system automatically uses inert gas for purgingsamplepipelineAnd quantitative loopTo prevent cross contamination;
5The heating position has oscillation function,Adjustable speed,Can reduce the time required for the sample to reach equilibrium;
6One click call to stored headspaceProgram, press the run button to automatically complete all sample analysis,Real time animation displaySample dynamics;
7The sample transfer tube and injection valve have automatic blowback function, which avoids cross contamination of different samples;
87-inch touch screen display, easy to operate,And display the dynamic flow path of the gas path in real-time through images;
9. Real time display of the number and location number of samples being heated in the heating position on the screen, as well as real-time display of the sample to be injected and the sample to be injected;
10. Using AFM technology, the carrier gas and pressurized gas can be digitally displayed with an accuracy of 0.01psi, ensuring consistent conditions for each headspace sampler and guaranteeing quantitative analysis results;
11The instrument is equipped with an electronic version for easy teaching, making it convenient for beginners to learn and use on the spot;
12. Compatible with various brands and models of GC both domestically and internationally, and can be connected in series;Not occupying the injection port, no need to remove the liquid automatic sampler; Parallel connection can also be used for convenienceComplete the replacement of different GCs within 30 seconds.
IIMain technical parametersrequirement
1Temperature control range of sample area: room temperature260℃,In increments1 ℃ arbitrary setting,Heating power approximately400W;
2Temperature control range of valve injection system: room temperature220℃, In increments1 ℃ arbitrary setting,Heating power approximately60W;
3Temperature control range of sample transfer pipeline:room temperature220℃,In increments1 ℃ arbitrary setting,Heating power approximately40W;
4Temperature control accuracy:≤ ±0.1℃;
5Headspace bottle workstation:45bitConvenient for storing sample bottles;
6HeatingworkPosition:15bit;
7Headspace bottle specifications:10ml, 20ml (standard) optional;
8. HeavyRestoration:RSD ≤ 1% (ethanol in 200ppm water, N=5);
2.9 Injection volume control mode:Quantitative ring injectionorPressure balanced injectionControl the injection volume;
2.10 injectionCarrier gas pressure setting:0~0.4MPA (continuously adjustable);
2.11 backblowGas pressure setting:0~0.4MPA (continuously adjustable).
3、 Configuration List
1. 45 positionsFully automatic headspace sampler host 1platform;
2Jaw headspace bottle 100 sets;
3Open lid pliers 1 piece;
4communication line 1 piece;
5Power cord, certificate of conformity, warranty card, instruction manual, etc1 set.