Thermo FisherTG-5MSGas chromatography column
(Item number)26098-1420)It is widely used inGC/MSA low polarity capillary column for analysis, suitable for high-sensitivity detection of residual solvents, environmental pollutants, and active compounds in complex matrices.
Product Name:TraceGOLD™ TG-5MS GCchromatographic column
Item No.26098-1420
Brand: Thermo Fisher Scientificthermo
Fixed phase:TraceGOLD™ TG-5 MS(5%phenyl-95%Dimethyl polysiloxane)
USPClassification:G27, G36
Product line:TraceGOLD
Length:30 m
Inner diameter:0.25 mm
Film thickness:0.25 µm
temperature range:-60°Cto325°C(Using temperature)upper limit)Recommended for long-term use not exceeding305°C
USPclassification:G27, G36
Applicable detector:FIDTheECDTheTSDTheGC/MS(Recommended)
application fieldResidual solvent analysis(ICH Q3C)Polycyclic aromatic hydrocarbons(PAHs)Pesticide residues and volatile organic compounds(VOCs)Petrochemical product analysis, etc
Suitable for: with signal-to-noise ratio, sensitivity, andMSIntegrity ofGC/MS, used for analyzing active compoundsGC/MS
Key performance advantages
Low loss designExclusive bonding and capping process significantly reduces column loss, improves mass spectrometry signal-to-noise ratio, and extends detector maintenance cycle
High inert surfaceSuitable for the analysis of compounds containing active functional groups (such as acids, alcohols, amines), reducing tailing and improving peak symmetry
Good batch consistencyEach column comes withCOA(Quality certificate), ensuring the repeatability of method transfer and verification
Strong compatibility: AdaptationAgilentTheShimadzuThePerkinElmerMainstream brands such as Thermo Fisher ScientificGC/GC-MSsystem
Precautions for use and maintenance
Installation directionBe sure to follow the labeling on the chromatographic column“→”Flow installation, connect the injection port end to the gasket for purging, and connect the detector end to the mass spectrometry ion source
Aging treatmentNew columns or columns that have not been used for a long time need to undergo programmed temperature aging
Suggested procedure:50°C(2 min)→ 10 °C/min → 320°C(30 min)
Carrier gas flow rate:1.0 mL/minNitrogen or helium,Do not heat the chromatography column without a carrier gas
Avoid pollution:
Ensure the sample is clean before injection, it is recommended to use0.22 μmmembrane filtration
Samples containing particles, strong acids, strong bases, or high boiling point substances require pre-treatment (such as...)SPELiquid liquid extraction
Cutting the column headIf there is a peak tailing or retention time drift, the injection port end can be cut off each time10–20 cmRe aging
Storage conditionsSeal both ends with sealing caps and store in a dry and cool place to avoid severe vibration