Particle filtration efficiency tester ORF-RH1890
Before introducing the testing instruments, first of all, popularize the following relevant knowledge:
1、 Filtration efficiency:The percentage of particulate matter filtered out by the filter element under specified testing conditions;
2、 Oil mist detection:Use oil-based or equivalent substances such as DEHP/PAO/DSP as aerosols to test the filtration performance of the hood or filter material;
Salt spray detection:The method of using NaCl or equivalent halides as aerosols to test the filtration performance of hood or filter media;
The PFT-01 particle filtration efficiency tester developed and produced by Shenzhen Ouruifan Equipment Co., Ltd. using oil mist and salt mist methods is used for testing the particle filtration efficiency of daily protective masks, medical masks, and respirators.
3、 Main parameters of particle filtration efficiency tester:
Other names: Particle Filtration Efficiency Tester, Medical Kou Mask Particle Filtration Efficiency Tester, etc
1. Efficiency detection range: 0-99.999%
2. Accuracy: Level 1
3. Flow meter range for efficiency detection: 10-100L/min
4. Flow meter accuracy: Level 2.5
5. Sampling frequency: 1-9999 times/min
6. Filtration efficiency: Particle concentration: 20-30mg/m3
7. Median diameter for counting: salt particles (0.075 ± 0.02) μ m, oil particles (0.185 ± 0.02) μ m
8. Geometric standard deviation of particle size distribution: salt particles ≤ 1.86, oil particles ≤ 1.60
9. Dynamic detection range: 0.001-100) mg/m3, accuracy 1%
10. Host power supply: 220VAC 50Hz
11. Net weight of the host: 64 kg
12. Instrument configuration: host, computer, professional software, aerosol generator, laser dust particle counter, glass rotor flowmeter, vacuum pump.
Particle filtration efficiency tester ORF-RH1890
The structure of this detection instrument includes: spray pump, electrostatic neutralizer, cylinder, flowmeter, PLC industrial control system, fan, exhaust gas filter, flow transmitter, differential pressure sensor, touch screen computer, oil mist aerosol generator, salt mist aerosol generator, etc;