- Phone
-
Address
Room 2202, International Building 1, Optics Valley Headquarters, No. 58 Optics Valley Avenue, Donghu New Technology Development Zone, Wuhan City
Wuhan Deke Tong Technology Co., Ltd
Room 2202, International Building 1, Optics Valley Headquarters, No. 58 Optics Valley Avenue, Donghu New Technology Development Zone, Wuhan City
Guwei LCR-6002/6020 2KHz safety tester The series also offers a variety of auxiliary measurement modes to meet the needs of different types of materials in measurement applications. For example, for multilayer ceramic capacitors(MLCC)Automatic level control is provided to meet the demand for testing voltage in measurement(ALC)Function. For the measurement of inductive components, it provides adjustable test current, DC resistance measurement function, and optional external bias current box(±2.5A)To meet the measurement needs. For the DC bias test required for capacitor components, internal ±2.5VAdjustable voltage, or through the optional external bias box(±45V)Allow users to perform material validation measurements. Besides,10The step list testing function allows the parameters of each testing step to be set according to the user's needs, in order to observe the trend of the characteristics of the tested component.
LCR-6000The series providesHandlerThe interface, coupled with its own sorting function for measurement(9BIN,AUX:1BIN)It can provide a connection sorting machine for material sorting operations. For users who require remote control and retrieval of measurement results,LCR-6000The series providesRS-232C, allowing the reading of control or measurement results to be set throughPCSend instructions to proceed; Additionally, it is freePCEnd software not only provides users with a tool to immediately store measurement results, but also saves time on program development.

LCR-6000Although the series has been fully miniaturized, the functions it provides are more diverse and abundant. Firstly, the entire series adopts3.5Inches of colorLCDDisplay screen, in addition to maintaining the same rich display parameters as before~In addition to displaying the setting conditions and measurement results simultaneously, it also adds2Additional monitoring parameters; In short, multiple displays can be made on the same screen4The parameter results greatly improve the efficiency of measurement. In addition, a large value display mode switch has been added, which not only emphasizes the measurement results, but also providesPASS/FAILThe judgment of detection makes it faster and more convenient.
Another feature of this series of products is that they are more convenient to use.LCR-6000The series provides two ways to perform zeroing~Full frequency band or single point; After the user performs full frequency zeroing, they can measure at any frequency point within the applicable frequency range without turning off the power or replacing the test fixture, saving time waste caused by repeated zeroing; In addition, the frequency within the range is continuous, and users can input the actual required frequency value to conduct real testing on the components.
LCR-6000Series product features:
3.5'colorfulLCDdisplay screen
5A model(10Hz~2kHz/20kHz/100kHz/200kHz/300kHz)
Test frequency continuously adjustable
Basic accuracy : 0.05%
Fast testing speed25mS
Can choose between full frequency band or single pointOPEN/SHORT
16Seed Owner/Secondary parameter measurement combinations are available for selection;Additionally, there are options available2An additional monitoring parameter(It can be displayed once more4Different parameter values)
provideDCRMeasurement and internal DC bias voltage(DC BIAS,±2.5V)
Fixed latitude LCR-6002/6020Quick selection
model test frequency Basic accuracy 10Step by step testing function USBStorage function Standard interface
LCR-6300 10Hz-300KHz 0.05% YES YES USB/RS-232
LCR-6200 10Hz-200KHz 0.05% YES YES USB/RS-232
LCR-6100 10Hz-100KHz 0.05% YES YES USB/RS-232
LCR-6020 10Hz-20KHz 0.05% YES YES USB/RS-232
LCR-6002 10Hz-2KHz 0.05% YES YES USB/RS-232
Detailed Specifications
test frequency
LCR-6300 10Hz~300kHz(±0.01%) (4Position display)
LCR-6200 10Hz~200kHz(±0.01%) (4Position display)
LCR-6100 10Hz~100kHz(±0.01%) (4Position display)
LCR-6020 10Hz~20kHz(±0.01%) (4Position display)
LCR-6002 10Hz~2kHz(±0.01%) (4Position display)
output impedance
30Ω/ 50Ω/ 100Ω can be selected
Basic accuracy
slow speed /medium speed 0.05%
fast 0.10%
test speed
fast: 25ms /medium speed: 100ms /slow speed: 333ms
Test signal level
voltage 10.00mV- 2.00V (±10%) CV:10.00mV- 2.00V(±6%)
current 100.0uA- 20.00mA (±10%) CC:100.0uA- 20.00mA(±6%) (@2VMax)
DC bias voltage
internal ±2.5V (0.5%+0.005V)
display range
R, X, |Z| 0.00001Ω~ 99.9999MΩ
G, B, |Y| 0.01nS ~ 999.999S
L 0.00001uH ~ 9999.99H
C 0.00001pF ~ 9999.99mF
D 0.00001 ~ 9.99999
Q 0.00001 ~ 99999.9
θd -179.999°~ 179.999°
θr -3.14159 ~ 3.14159
DCR 0.00001Ω~ 99.9999MΩ
Δ% -99999% ~ 99999%
test parameters
test combination Cs-Rs, Cs-D, Cp-Rp, Cp-D, Lp-Rp, Lp-Q, Ls-Rs, Ls-Q,Rs-Q, Rp-Q, R-X, DCR, Z-θr, Z-θd, Z-D, Z-Q, Auto LCZ
monitoring parameters(most2A selectable option/each time) Z, D, Q, Vac, Iac,Δ,Δ%,θr,θd, R, X, G, B, Y
Table testing
10 Step by step
sorting
comparator (9BIN,AUX:1BIN)
Memory group
internal - Panel settings 10A file name
internal - Measurement data 10000 Data(.csv)
USBstore 10A file name, nine thousand nine hundred and ninety-nineaDataarchive, nine hundred and ninety-nineaLCDPicture Archive
Other functions
Automatic positioning control (ALC) ON/OFF
average 1~256time
trigger internal /manual/外部/ Bus
Trigger delay 0ms~60s
Judgment function PASS / FAIL
Image storage Direct depositUSB(BmpFormat)
display screen
3.5inchLCD, RGB color (320x240)
interface
RS-232(SCPI), Handler, USB Host
Use power supply
AC 90V-250V, 50-60Hz, Max. 30VA
Dimensions and Weight
265(wide)x107(tall)x312(long) mm,about3kilogram