- Phone
-
Address
Room 9011, Building B, Lane 519, Macau Road, Putuo District, Shanghai
Shanghai Xudun Electronic Technology Co., Ltd
Room 9011, Building B, Lane 519, Macau Road, Putuo District, Shanghai
+GF+Signet low conductivity pH pure water electrode
3-2726-LC-00 159 001 557 electrode, pH, Light bulb type glass, low conductivity, PT1000, 3/4 in. NPT thread*
PH system calibration
How to Calculate the Slope Efficiency of pH Electrode
Select two pH buffer solutions to check the performance of the pH probe, pH 4.01 and pH 7.00. The ratio of the measured potential (E2-E1) to the pH value (7.00-4.01) gives the slope of the line.
1. Read the mV potential generated in two calibration buffers.
2. Determine the slope (mV potential generated per pH unit).
3. Divide this number by the theoretical slope, pH 59.16 mV/@ 25 ° C, and multiply by 100.
Example 1
The pH electrode generates -12 mV in pH 7.00 buffer solution, at pH
+162 mV is generated in buffer solution 4.01.
160 mV -(-12 mV)=172 mV
172 mV/3 =57.33 mV/pH unit
57.33/59.16 x 100=96.9% efficiency
Example 2
PH electrode generates -45 mV in pH 7.00 buffer solution, pH
+115mV in buffer solution 4.01.
115 mV -(-45 mV)=160 mV
160 mV/3 =53.33 mV/pH unit
53.33/59 .16 x 100 =90.1%
In Example 2, a slope efficiency of 90.1% is acceptable, but an offset of -45 mV is not acceptable. Offset is possible It is the result of contamination or poisoning of the reference electrode. Slope offset
It may be the result of glass being etched or adhering to hard objects.
As the electrode ages, the slope gradually decreases, and a slope of 85% -105% is generally acceptable. If the bias value (pH 7.00) is>+/-45mv, the electrode should be replaced. When the electrode needs to
When seeking high accuracy, the offset and slope values need to be as close as possible to the theoretical values.
Please refer to the calibration section of the instrument manual for 2-point pH calibration. If it is a 4-20mA non display application, please refer to the calibration section of 2751 Smart Sensor Signal Amplifier for further instructions
Two point pH calibration.
Due to the constantly changing characteristics of pH electrodes, pH electrode calibration is an important task that must be frequently carried out in electrode standardization work. The electrode contains a gel
The instrument will be depleted, so it must be regularly readjusted to maintain the accuracy of the system. There are many factors that affect the performance of probes, including aging, temperature, coating, and chemicals. in
After cleaning and adjustment, the Signet pH electrode can be calibrated in two pH buffer solutions with different pH values (two-point calibration).
The pH electrode is calibrated by placing it in a buffer solution with a known pH and measuring the battery potential. The pH value is a linear function within the pH range of 2-11
Number, therefore two calibration points are required. The most common calibration buffers are 4.01 and 7.00 (@ 25 ° C). If the pH value is measured in the acidic to neutral range,
We recommend using buffer pH 4.01 and buffer pH 7.00 for calibration. If measuring alkaline samples, we recommend using buffer 7.00 and buffer
Perform calibration on 10.01.
The response of glass electrodes is mainly composed of two parameters: offset and slope. The theoretical output of the pH electrode is 0 when the pH is 7
MV. Offset is the deviation of pH value from the nominal value. In Signet transmitters, offset calibration is performed under the standard.
The slope of the glass electrode is the mV generated per pH unit (mV/pH). At 25 ° C, the theoretical slope is 59.16 mV/pH. Therefore, at pH 4.01
At pH 10.01 (with lower H+concentration), a potential of -177 mV will be generated (see Table 2). For a new electricity
In extreme terms, at pH 7.00, 0-15 mV will be generated, with a slope efficiency ranging from 93.2% to 103%. In Signet
In the transmitter, the manual slope calibration operation menu is located in SLOPE.
+GF+Signet low conductivity pH pure water electrode
description
The Signet2724-2726 and 2734-2736 series pH/ORP electrode products are designed to reduce maintenance time and enable
Using value.
• Unique DryLoc ® The connector provides quick assembly and reliable connection, including gold-plated contact surfaces and O-ring seals.
The design has extended the lifespan of the electrode by lengthening the path inside the reference chamber.
The selection of liquid receiving materials can withstand various industrial applications.
Multiple installation methods address the need for installation flexibility. Fix on the pipe using NPT and ISO7/1 R-inch threads
On the road, immerse it in a tank or directly install it on a similar installation accessory of SIGNET.
• Pair the 2724-2726 or 2734-2736 electrodes with the Signet 2751 smart sensor signal amplifier for programmability
Logic controllers (PLCs), SCADA systems, or data acquisition programs provide 4-20mA loop signals.
Storage chips can store production, manufacturing, calibration, and operational data.
-Production and manufacturing information (serial number, model, and factory glass impedance).
-Calibration information (factory calibrated pH/ORP, factory calibrated temperature, user calibrated pH/ORP, user calibrated)
Temperature, offset pH/ORP, and slope% pH/ORP.
-Operating data (sensor operating time, minimum and maximum measured pH/ORP, minimum and maximum measured temperature)
To achieve greater flexibility and unique features, please pair the 2724-2726 or 2734-2736 series with the 2751 Smart
Sensor electronics paired with 9900 transmitter, 9950 dual channel transmitter, or 0486 Profibus hub
Use it. Other features include electrode health monitoring through glass impedance measurement, glass breakage detection, and reference impedance measurement
Testing and diagnosis. Convenient remote calibration eliminates the hassle of on-site calibration.
Using digital (S ³ L) signal output, connect six 2724-2726 or 2734-2736 electrodes to an 8900 multiparameter
On the controller, use 6 2751 pH/ORP sensor signal amplifiers
Using digital (S ³ L) signal output, pair two 2724-2726 or 2734-2736 electrodes with two 2751 pH/ORP
The sensor signal amplifier is connected to a 9950 dual channel transmitter.
The electrodes 2724-2726 are recommended for general use applications.
2734-2736 electrodes are recommended for harsh applications.
The 2734-2736 electrode utilizes advanced reference technology to extend its lifespan in almost all applications.

Technical Specifications
General
Compatibility 2751 pH/ORP signal amplifier
2760 pH/ORP preamplifier (applicable only)
2724-2726 3K Balco 型)
PH temperature sensor:
PT1000 and 2751 signal amplifiers are compatible
Connect to PLC, 9900 transmitter,
8900 multi parameter controller,
9950 dual channel transmitter
3 K Ω Balco compatible with 2760 preamplifier
Connect to ........................ 8750 transmitter
Process connection 3/4NPT ISO7/1R3/4 thread or SIGNET fittings
wetted material
PH (2724&2726) polyphenylene sulfide, glass, ultra-high molecular weight polyethylene,
fluororubber
ORP (2725) polyphenylene sulfide, glass, ultra-high molecular weight polyethylene,
Fluororubber, platinum
PH (2734&2736) polyphenylene sulfide, glass, PTFE, fluororubber
ORP (2735) polyphenylene sulfide, glass, PTFE, Fluororubber, platinum
Weight .....................................0.25 kg (0.55 lb)
performance
efficiency
2724-2726 >97%@25 °C (77 °F)
2734-2736 >95%@25 °C (77 °F)
Efficiency reflects the "health" indicator of the new electrode
Efficiency is the actual slope value (mV/pH) at 25 ° C
Ratio to theoretical value of 59.16 mV/PH
95% to 100% efficiency, equivalent to a slope between 56.20 and
59.16mV /pH
measurement range
pH.........................................0 to 14
ORP. .....................................±2000 mV
3-2726-LC Low Conductivity Fluid (20 to 100 μ S/cm)
≤ 20 µ S/cm...... The flow rate should be controlled at 150 ml/min
A reliable and grounded system
3-2724-HF,3-2726-HF,3-2734-HF,3-2736-HF:
Glass resistant, with a pH value of 6 or below; Trace HF ≤ 2%
In applications with a concentration of 2% or lower, ordinary pH glass is used at a pH of 6 or below
They will all be eroded.
Environmental Requirements
Operating Temperature
2724-2726* -10 °C to 85 °C (14 °F to 185 °F)
2734-2736* 10 °C to 100 °C (50 °F to 212 °F)
**3-2724-HF,3-2726-HF,3-2734-HF,3-2736-HF 温度高
Best performance at 10 ° C (50 ° F)
