+GF+Signet Turbidity Meter 0-200NTU in Large Stock
+Application of GF+Signet Turbidity Meter
Water plant inlet and outlet
End of pipe network water
secondary water supply
Pool water and membrane process water treatment equipment (direct drinking water, tap water, sewage)
System Features
Through integrated design, users can reduce the construction work of turbidity sensor water circuits by simply connecting one inlet pipe and one outlet pipe for measurement.
The measuring tank is equipped with a defoaming design to ensure the accuracy of turbidity measurement.
View and control through a transmitter, RS485 data output, convenient for integration and control with the central control room or upper computer, and configured with 4-20mA output
Output and relay control to meet the needs of various on-site control applications.
The sensor is plug and play, easy to install and maintain.
Intelligent sewage discharge function reduces the frequency of manual maintenance. Product Introduction
The iTurb 4160 turbidity online analyzer combines digital turbidity sensors with intelligent technology
The turbidity online analyzer and waterway can be integrated into a complete machine. In turbidity analysis
On the analyzer, measurement data can be viewed and managed, and calibration and other operations can be performed. that
The system integrates online analysis of water quality turbidity, remote data transmission, and data storage
Integrated with calibration function, it provides data collection and analysis for water turbidity
Extremely convenient.
The turbidity sensor measurement tank is equipped with a defoaming design, which can make the water sample after defoaming
Enter the turbidity measurement tank again. This instrument requires fewer water samples and has strong real-time performance.
A fine water sample flows through the defoaming grid and enters the measuring tank, and is always maintained
The water sample is circulated, and the instrument collects the turbidity of the water sample when it passes through the circulation device,
It can also be connected to the central control room or upper computer through digital communication.
+GF+Signet Turbidity Meter 0-200NTU in Large Stock
Technical specifications of iTurb 4160 turbidity meter 0-200 NTU

Parameter technical specifications
measurement parameters
turbidity
Measuring principle
90 ° scattered light method
The light source is a laser
Compliant with EPA 180.1 standard
technical parameters
range
Turbidity: 0.2~200NTU
Check it out
0.2NTU
accuracy
± 5% or ± 0.5NTU, whichever is greater
zero drift
± 0. 5NTU
display resolution
Turbidity: 0.1NTU
Measurement interval
continuous
Shell&Display
display screen
240 * 160 dot matrix LCD screen
material
The header and cabinet are both made of PC plastic
Overall dimensions
Length: 400mm Width: 230mm Height: 600mm
Protection level
IP65
Software features
Function password management
Function password: set 3700, calibrate 3900, historical data 1300, historical curve 1400
historical data
Yes, the storage interval is 5 minutes
Data Logger
With data recording, it can store 1000 pieces of data
language
Chinese/English
output
serial port output
1 RS485, Modbus RTU protocol
current output
2-channel 4-20mA output
relay output
3-channel relay normally open contact output, timed discharge)
Maximum load: 5A/250VAC or 5A/30VDC
Installation requirements
Installation location
Indoor wall mounted
Imported pipe diameter
6mm
出口管径
10mm
power supply
Supply voltage 100-240VAC, 50/60Hz
Power<20W
Requirements for measuring water quality
< 200 NTU
influent flow rate
<250mL/min
sample temperature
0~45°C
ambient temperature
0°C~50°C
Storage temperature
-40°C~70°C
humidity
<95% (uncondensed)

Installation and Wiring
Installation requirements
1) The installation location of the instrument should be chosen in a spacious, ventilated, and dry indoor environment;
2) The instrument should be placed and fixed at a position slightly above the drainage water level (300mm above ground level is the most suitable height) to facilitate smooth drainage;
3) Adequate space should be left around the instrument for personnel operation to facilitate later maintenance and calibration;
4) The horizontal height of the drainage pipe of the instrument should be lower than the horizontal height of the outlet of the instrument, and the total length of the outlet pipe should not exceed 3 meters, and there should be no blockage at the end.
5) The inlet and outlet pipes of the instrument are equipped with quick connectors, and the inlet pipes have matching conversion joints. Users can choose other conversion joints according to the actual situation on site, but
The pipeline ultimately connected to the instrument inlet cannot be changed;
6) Incoming water pressure: ≤ 0.3 Mpa;
7) Flow requirement: ≤ 250 mL/min;
8) Anti interference prevention measures: Do not share the same power supply with electric motors, frequency converters, and other high-power equipment;
9) Water and electricity measurement: After the instrument is powered on, open the inlet valve and the water sample enters the circulation pool. After the water level in the circulation pool reaches a certain height, the water sample flows out from the drainage outlet.