Welcome Customer !

Membership

instrumentb2bThe use of conductivity electrodes
  Conductivity electrodeApplicable fields: It can be widely used for continuous monitoring of conductivity values in solutions such as thermal power, chemical fertilizers, metallurgy, environmental protection, pharmaceuticals, biochemistry, food, and tap water.
  conductivity meterThe use of
1. Adjust the conductivity meter: Before turning on the power, observe whether the needle points to zero. If it does not point to zero, adjust the screw on the meter head to zero.
2. Turn the "Calibration, Measurement" switch to "Calibration".
3. Connect the power supply and preheat for more than 10 minutes (until the pointer * stabilizes). Turn the range switch to X 104 and then use the "Adjust" knob to adjust the pointer to 1.0 full degree
4. Measure the conductivity value of solvent water: Add 50ml of water to a 150ml conical flask and immerse it in the electrode.
5. Rotate the "constant" knob to the corresponding value based on the conductivity constant on the electrode.
6. Turn the "Range" switch to × 102, set the conductivity meter to "Low Cycle", and then use the "Correct" knob to adjust the pointer to 1.0 full degree.
7. Read the conductivity value of solvent water.
8. After the measurement is completed, turn the "calibration, measurement" switch to "calibration".
  Conductivity electrodePrecautions for use
1. The electrode plug socket should not be affected by moisture, and the instrument should be placed in a dry environment to avoid measurement errors or leakage caused by moisture or water droplet splashing.
To avoid drift in measurement values when measuring pure or ultrapure water, it is recommended to use a sealed groove for flow measurement in a sealed state. If a beaker is used for sampling and measurement, significant errors will occur and the measurement will be inaccurate.
3. As the temperature compensation system adopts a fixed 2% temperature coefficient compensation, it is recommended to measure high-purity water without compensation as much as possible. After measurement, check the table for the situation.
4. The electrode should undergo regular constant calibration.
5. To ensure measurement accuracy, the electrode should be rinsed twice with distilled water (or deionized water) less than 0.5 μ S/cm before use (the platinum black electrode should be soaked in distilled water for a while before use), and then rinsed three times with the test sample before measurement.
6. The measuring electrode is a precision component that cannot be changed in shape and size, cannot be decomposed, and cannot be cleaned with strong acid or alkali to avoid changing the electrode constant and affecting the accuracy of instrument measurement.
9. Other solutions are as above.
Latest News
  • Important Notice: Notice on Organizing Participation in the 2021 Shenzhen International Sensor and Intelligent Instrument Exhibition

    Important Notice: Notice on Organizing Participation in the 2021 Shenzhen International Sensor and Intelligent Instrument Exhibition

    The "2021 Shenzhen International Sensor and Intelligent Instrument Exhibition" jointly organized by the Shenzhen Sensor and Intelligent Instrument Industry Association, Guangdong Zhizhan Exhibition Co., Ltd., and Guangzhou Zhizhan Exhibition Service Co., Ltd. was held at the Shenzhen Convention and Exhibition Center from July 1-3, 2021. $r$n
    01,21,2026 76 views
  • Big broadcast! Search for concurrent conferences at the Munich Shanghai Light Expo, just read this article!

    Big broadcast! Search for concurrent conferences at the Munich Shanghai Light Expo, just read this article!

    The 16th Munich Shanghai Light Expo will be held from March 17-19, 2021 at the Shanghai New International Expo Center. The registration channels for concurrent conferences have now been fully opened. The conference has a wide range of topics, focusing on industry hot topics, including laser technology, advanced lasers, fiber lasers, optical technology, optical detection, intelligent manufacturing, infrared imaging technology, laser safety, LiDAR, autonomous driving and other new research and application achievements, covering a wide range of laser and optoelectronic fields. In the warm spring of March, we will continue to bring you more exciting topics! $r$n
    01,21,2026 122 views