Constant temperature and humidity incubators are capable of conducting tests such as high temperature, low temperature, and humid heat. If various temperature and humidity faults occur during the testing process, the following methods can be used to handle them:
1. The low temperature of the constant temperature and humidity incubator does not meet the set indicators in the experiment
It is necessary to observe whether there is a change in temperature, whether the temperature drops slowly or if there is a tendency for the temperature to rise after reaching a certain temperature value.
The former must be checked for the reason, whether the workshop has been dried before conducting the low-temperature test, and after keeping the workshop dry, the test sample should be placed in the workshop for further testing. Then check if there are too many test samples placed in the studio, causing the air inside the studio to not circulate. After ruling out the above reasons, it is necessary to consider whether the refrigeration system of the incubator has malfunctioned. If so, professional repair personnel from the supplier should be hired for maintenance.
The latter phenomenon is caused by the constant temperature and humidity incubator being placed in a poor environment. The temperature and location of the constant humidity incubator (the distance between the back of the box and the wall) must meet the requirements