The main calibration steps for oscillating water bath pots
Date: 2025-06-12Read: 0
Oscillating water bathMainly suitable for oscillation cultivation of various liquid and solid compounds such as biology, biochemistry, cells, and bacterial strains in various universities, medical, petrochemical, health and epidemic prevention, environmental monitoring and other scientific research departments. This product has the characteristics of reasonable structure, easy operation, and high stability performance, making it an ideal equipment for laboratory workers to handle with ease.
Product Features:
1. Temperature control range: -5-100 degrees (with cooling).
2. Temperature control method: Adopting microcomputer temperature control, precise temperature control, no overheating or overshoot phenomenon.
3. There are three oscillation modes: rotary, reciprocating, and dual function. (Choose when placing an order)
4. Inner liner material: high-quality stainless steel, universal spring clamp, and different clamps can also be designed according to requirements.
5. Shell material: High quality cold plate surface with electrostatic spraying, rust prevention, and extended service life.
6. Motor type: Permanent magnet DC motor, small size, high power, long service life, and also has the characteristics of low noise.
7. Speed regulation mode: stepless speed regulation, or microcomputer closed-loop control speed regulation mode (speed display, speed deviation of 0) can be selected.
8. Safety category: equipped with leakage protection, automatic power-off function, over temperature protection, and sound and light alarm.
9. Timing range: 0-120 minutes or microcomputer 0-9999 minutes with timing and normally open function.
Core verification steps:
Project: Temperature deviation; Operation method: Set the target temperature (such as 37 ℃, 60 ℃) → stabilize for 30 minutes → record the deviation between the measured value at the center point and the set value.
Project: Temperature uniformity; Operation method: Install ≥ 5 sensors (center+four corners) → Calculate the highest/lowest temperature difference.
Project: Temperature stability; Operation method: Record the temperature every 2 minutes → Continue for 30 minutes → Calculate the fluctuation range.
Project: Insulation performance; Operation method: Use a 500V insulation resistance meter to test the resistance between the energized part and the casing (≥ 2M Ω).