How to deal with the auxiliary system failure of the cell smear centrifuge
Date: 2025-09-11Read: 1
Cell smear centrifugeWidely applicable to industries such as teaching, scientific research, biochemical products, medical inspection, food safety, agricultural product testing, animal husbandry and aquatic product testing, microbiology, molecular chemistry, environmental protection, quality inspection, etc. Product large screen true color LCD display screen, full touch screen operation interface; Microcomputer control, brushless AC variable frequency motor drive, high speed accuracy, low noise; Programmable operation, multiple speed options for lifting and lowering; It can independently set centrifugal force and speed and display them on the same screen.
Auxiliary system malfunction:
Display screen abnormality:
Possible reasons: damage to the control board or sensor malfunction.
Solution: Replace the display module or calibrate the sensor.
The door cover cannot be closed:
Possible reasons: aging of sealing ring or malfunction of door lock motor.
Solution: Replace the sealing ring or inspect the door lock device.
Motor and starting faults:
Unable to start:
Possible reasons: power failure, blown fuse, worn carbon brush, or open circuit in motor coil.
Solution: Check the status of the power cord and fuse, replace the carbon brush or repair the motor circuit.
Unstable rotational speed:
Possible reasons: voltage fluctuations, sensor malfunctions, or rotor imbalance.
Solution: Use a voltage regulator to stabilize the voltage, calibrate the speed sensor, and symmetrically place the centrifuge tube.
Precautions for use:
Before using the centrifuge, it must be placed on a stable and sturdy ground (tabletop).
2. After use, it should be allowed to stop running on its own, and it is strictly prohibited to open the hood while it is still running or when it is turned on.
3. If abnormal sound is found during use, the device should be turned off immediately and checked and repaired.
4. Use the centrifuge strictly according to the operating procedures.
5. It is not allowed to use beyond the maximum number of revolutions of the rotor.