- Phone
-
Address
Tongxie Jinzuo, 977 Huafeng Middle Road, Shangcheng District, Hangzhou City
Hangzhou Botao Purification Equipment Engineering Co., Ltd
Tongxie Jinzuo, 977 Huafeng Middle Road, Shangcheng District, Hangzhou City
biosafety cabinetofoperationmatter
1Before operation, all items required for this operation should be moved into the safety cabinet to avoidFrequently crossing the air curtain with both arms to disrupt the airflow; And disinfect the surface with 70% alcohol before moving in to remove contamination.
2. Turn on the fan for 5-10 minutes, wait for the air inside the cabinet to be purified and the airflow to stabilize before conducting the experimental operation. Slowly insert both arms into the safety cabinet and let it rest for at least 1 minute to stabilize the airflow inside before proceeding with the operation.
3. No items unrelated to this experiment shall be placed in the safety cabinet. The placement of items in the cabinet should ensure that the clean area, semi polluted area, and polluted area are basically separated, and the items are easy to access during operation, with no crossing between the three areas. Items should be placed as far back as possible, but should not block the airway to avoid interfering with the normal flow of air.
4. The operation should be carried out from the clean area to the contaminated area to avoid cross contamination. To prevent possible splashing droplets, a towel or gauze soaked in disinfectant can be placed on the countertop, but it cannot cover the safety cabinet grille.
5. During the operation inside the cabinet, it is strictly prohibited to use open flames such as alcohol lamps to avoid the heat generated generating airflow and interfering with the stability of the airflow inside the cabinet; And open flames may damage HEPA filters.
6. Try to minimize the movement of people behind and quickly open and close doors during work to prevent unstable airflow inside the safety cabinet.
7. During experimental operations, the glass window should not be opened and the operator's face should be kept above the work window. When operating inside the cabinet, the movements should be gentle and soothing to prevent affecting the airflow inside the cabinet.
8. The safety cabinet should be regularly inspected and maintained to ensure its normal operation. Once abnormal operation of the safety cabinet is discovered during work, work should be immediately stopped, corresponding measures should be taken, and relevant personnel should be notified.
9. After work, close the glass window and keep the fan running for 10-15 minutes. At the same time, turn on the ultraviolet light and irradiate for 30 minutes.
10. The safety cabinet should be cleaned and disinfected regularly. The contaminants on the countertop of the cabinet can be wiped with 2% 84 disinfectant after the work is completed and disinfected with ultraviolet light. The outer surface of the cabinet should be wiped daily with 1% 84 disinfectant.
11. Items used in the cabinet should be disinfected before being removed to prevent contamination of the environment by pathogenic microorganisms.
biosafety cabinetPrecautions for:
1. Slow movement principle: In order to avoid affecting the normal air flow state, hands should move as gently as possible when operating inside the cabinet.
2. Principle of parallel placement of items: In order to avoid cross contamination between items, items placed in the cabinet should be arranged horizontally in a straight line as much as possible to avoid cross contamination during the return air process. At the same time, avoid blocking the back return air grille and affecting the normal airflow path.
3. Principle of avoiding vibration: Try to avoid the use of vibrating instruments (such as centrifuges, vortex oscillators, etc.) inside the cabinet, as vibration can cause the particles accumulated on the filter membrane to shake off, resulting in a decrease in the cleanliness of the operating room. At the same time, if the balance fails on the front operating surface, it can also cause contamination of the safety cabinet to the operator.
4. Principle of moving different sample cabinets: When two or more items need to be moved inside the cabinet, the principle of moving low polluting items to high polluting items must be followed to avoid large-scale pollution of the interior of the cabinet caused by high polluting items during the movement process.
5. Principle of using open flames: Try not to use open flames inside the cabinet. Because the small particles and impurities generated during the use of open flames will be carried into the filter membrane area, these high-temperature impurities will damage the filter membrane. When it is unavoidable to use it, it is advisable to use a low flame Bunsen burner.