ZDCY-80 Intelligent DC Low Voltage Resistance Tester
1general narrate, recount, state
ZDCY-80Intelligent DC Low Voltage Resistance TesterComposed of an intelligent DC low resistance tester and a DC resistance testing fixture. LCD displays various parameters and measurement results. The testing fixture has a potential fixture and a current fixture. Mainly used to test the DC resistance of winding wires. This instrument also comes with a temperature sensor,And directly display the temperature value on the display screen,At the same time, it has the function of automatic temperature coefficient compensation, and the measurement result obtained is the corrected standard20The resistance value of the sample in degrees,And it can automatically convert the resistance value into electrical resistivity. This instrument is also randomly equipped with a standard resistance box(10 mOh,100 mOh,1(Ω) One. For users to test and calibrate the instrument. The test clip can be used to measure the resistance of other conductors, and the test data can be uploaded to our company's independently developed quality management system software for statistical analysis and processing of the test data.
2Main technical parameters
2.1Measurement range:10μ Omega20KΩ
2.2Resolution:1μ Omega
2.3Accuracy:(20±5℃)≤±(0.1%+3One word)
2.4Distance between two potential fixtures:1000±1mm
2.5Distance between current clamps and adjacent potential clamps(50±)1mm
2.6Fixture blade width:2±0.2mm
2.7Adapt to sample diameter:0~18mm
2.8Environmental conditions for use: Temperature:0~40℃ (To ensure measurement accuracy, it is recommended to use a temperature of20±5℃) Relative humidity:<80%
2.9Input power supply:AC220V±10% 50Hz
2.10External dimensions:L×W×H mm 1350×260×300
2.11Total weight of the switchboard:27Kg
ZDCY-80 Intelligent DC Low Voltage Resistance Tester
1Operation and usage methods
3.1Plug in the power plug, the power socket must use a three hole power socket, and follow the national electrician installation standard, with the left socketNConnect neutral wire, right socketLThe grounding wire and center socket must be properly connected to the grounding wire of the unit power supply system.
3.2The four wires that are led out from the bottom threading hole on the front of the instrumentrootTest cable connectionConnected to the front panel of this instrumentcorresponding HcurTheHpotTheLpotTheLcur Four testing terminalsTighten the socket and tighten the nut.
3.3Turn on the power switch,LCD display:

preheat15 minutes.
Among them“R:”To measure the resistance value, “P:”To measure the resistivity, “T:”To measure the temperature value.
pressShortcut menu key“MEAS”, can enter the settings interface, as shown in the following figure:

“temperature compensation:copper”Compensate according to the temperature coefficient of copper,Press the adjacent soft key to change to aluminum or turn off temperature compensation
“Temperature setting”Press the adjacent soft key to correct the measured temperature value
“Electrical resistivity setting”Press the adjacent soft key to turn on or off the resistivity display function,Set line type and external dimensions
“Measurement settings”Press the adjacent soft key to select the measurement mode as direct reading mode or comparison mode
“short circuit”Press the adjacent soft key during zeroing to reset
The "line type" can be a circular or flat line, which can be changed within the measurement settings
The "diameter" can be moved with the cursor using arrow keys and entered using numeric keys
The "Machine", "Class", "Model", and "Inspection" can be moved with the cursor using arrow keys and entered using numeric keys. They are mainly used in conjunction with our company's quality management system software
3.4Zeroing: The randomly configured "measuring clip" is used to measure the resistance of other objects and for "zeroing". When resetting, insert the "measuring clip" into the measuring instrument panelHcurTheHpotTheLpotTheLcur Four testing terminalsInside the socketAnd short-circuit the measuring clip according to the relevant provisions in the "Operation Manual". Press againSoft key next to 'short circuit':
The instrument automatically resets the seven measurement gears.
If the measuring clip is not properly short circuitedAfter pressing the short circuit button, the LCD screen will display:
(hint:The test clip did not short-circuit as required)
Zeroing must be completedRemove the 'measuring clip', otherwise it will affect the accuracy of the test results. Before testing each batch of samples, the measuring instrument should be reset to zero.
3.5Instrument settings
3.5.1As this instrument is mainly used to measure the resistance of winding wires, it has been set to direct reading mode for measurement before leaving the factory, which means the measurement result is1 mThe resistance of long winding wires.
3.5.2Setting of environmental temperature
according toGB/T4074.5-2008Winding wire test method5: Resistance,Regulation: Resistance is20At ℃1m
DC resistance of long winding wire. When it's not there20When measuring at ℃, the temperature coefficient must be corrected. This instrument is designed according to standard specifications15℃~25The temperature coefficient within the temperature range of ℃ automatically corrects the measured resistance value. To ensure the accuracy of the measurement results, it is recommended to15℃ to25The measurement is taken at an ambient temperature of ℃, and the resulting measurement is the corrected value at the current temperature1mThe resistance value of the long sample does not require manual correction.
3.5.3Test conductor material setting: This instrument can detect copper(Cu)And aluminum(Al)Set up two materials, the steps for setting up are as follows:
As shown in the above figure:“temperature compensation:copper”Compensate according to the temperature coefficient of copper,Press the adjacent soft key to change to aluminum or turn off temperature compensation
After setting up the instrumentWhen the operating environment temperature and the tested material are not changed, there is no need to perform instrument settings according to section 3.5.

3.6Measure the sample
Attention: Before use, the oil hole should be filled with oil④ Inject an appropriate amount of engine oil into the sample to make the handwheel rotate flexibly during straightening and avoid wear on the screw ⑦.
First, press the positions of the two fixtures on both sidesRemove the paint film of the sample clamped under the potential and current fixtures ③ on both sides.Lift up the handles of the potential and current clamps on both sides② Open the potential and current clamps.When the sample is less thanWhen 3mm, a straightening device can be used to straighten the sample;Press one end of the sample onto the left clamp of the sample straightening device① Press the other end under the right clamp ⑥. Rotate the sample straightening handwheel ⑦ clockwise to straighten the sample. Press down the potential and current clamp handles ② on both sides to clamp the test sample tightly. The instrument automatically selects the gear based on the resistance of the tested sample and displays the measurement results. Namely: On the LCD screenRx=******, which is the corrected DC resistance value of the one meter long test sample at the current ambient temperature.
3.7After the test is completed, the handles of the potential and current fixtures on both sides must be lifted first② Open the potential and current clamps. Then lift the left and right clamp handles of the straightening device and remove the sample. To avoid damage to the potential and current fixtures due to the tension of the sample.
3.8Accuracy check
The measurement accuracy of this instrument has been calibrated in detail before leaving the factory, but it is possible that the measurement accuracy may decrease due to user error or data accumulation caused by long-term use. Therefore, based on the characteristic that users mainly use this instrument to measure the resistance of copper rods or winding wires, it is randomly equipped with10mOh,100mΩ and1Oh.3Standard Resistance Box1One. After a period of use, users can use a standard resistance box to perform accuracy checks on the instrument according to the following methods.
3.8.21Environment during inspection
Temperature:20±1℃
Requirements for the working environment of other reference instruments.
3.8.2Resistance accuracy inspection steps
3.8.2.1Remove the sample measurement line from the panel and insert the testing clip.
Note: The instrument must first be turned on and preheatedMore than 15 minutes.
3.8.2.2According to the instructions in this manual3.5.2to3.5.3Article will“temperature compensation:copper”Set to closed state
3.8.2.3 Clamp the randomly configured 10 m Ω, 100 m Ω, 1 Ω, and three standard resistors one by one onto the measuring clamp (note: the red clamp faces in the same direction) for measurement, and record the actual measured values.
3.8.2.4Compare the measured value with the standard resistance, and the error should be within1%Within. If there is a significant deviation, please contact our company.
Non professionals are not allowed to make arbitrary settings to avoid affecting the measurement accuracy of the instrument.