Yiou brand SENG-3680T dust particle counter 28.3L

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Yiou brand SENG-3680T dust particle counter 28.3L laser dust particle counter is used to measure the particle size distribution and number of dust particles per unit volume in clean environments, and can also detect cleanliness levels. Widely used in industries such as electronics, pharmaceuticals, semiconductors, optics, precision machining, food manufacturing, cosmetics manufacturing, medicine and health, biological products, hospitals, environmental protection, inspection institutes, aerospace and other production enterprises and research departments, as well as epidemic prevention stations, disease control centers, blood centers, quality supervision institutes and other institutions

Product Details

Yiou brand SENG-3680T dust particle counter 28.3L

Overview:

laser particle counterUsed to measure dust particles per unit volume in clean environmentsparticle size distributionAnd the number, Cleanliness level can also be tested.Widely used in the electronics industry, pharmaceutical industry, semiconductor industry, optics, and precision machining Fields, food manufacturing, cosmetics manufacturing Production enterprises and research departments in industries such as medicine and health, biological products, hospitals, environmental protection, inspection institutes, aerospace, as well as institutions such as epidemic prevention stations, disease control centers, blood centers, and quality supervision institutes.

Yiou brand SENG-3680T dust particle counter 28.3L

Standard:
Reference standards: JJF1190-2008、JJG846-2015、GB/T6167-2007

Cleanliness assessment criteria: National standardsISO, 209E, GMP (dynamic/static)


Instrument Features

7-inch touch screen display control operation

sampling flow28.3L/min

Level 100~30000 testing

UCL calculation (with 95% confidence) can directly display particle concentration (particles/cubic meter)

UCL Report Standard Quick Selection

AC/DC configuration is optional

Intelligent data management,20000 sets of data loop storage function

Temperature and humidity detection, automatic alarm

Sampling dataUCL data partition storage, independent query, concise and clear

Long life laser light source with stable performance

Built in high-speed thermal printer

Printing options available, automatic instant printing, manual selection of printing function

The DC type has a built-in large capacity lithium-ion battery, imported battery cells, and an intelligent power management module that is more energy-efficient and safe. The split and detachable design is convenient for transportation and carrying.

Technical parameters:

sampling flow

28.3L/min(1.0ft3)

particle size channel

6通道 0.3 μm、0.5μm、1.0μm、3.0μm、5.0μm,10.0um

data storage

20000 sets of data for cyclic storage, querying, and power-off storage

sampling period

5~999999 seconds self set

test method

Single, continuous

working hours

≈ 8 hours of continuous DC testing

Charging time

≈ 4-hour DC type

UCL sampling frequency

2~50 self setting

Threshold alarm

Individual channel and any combination channel threshold alarm

repeatability error

superior toJJF1190-2008 standard requires ≤ 10% FS

Printing method

Built in high-speed thermal printer

light source

High precision laser sensor

Display Operation

7.0-inch touch screen

Verification standards

GB / T 6167 - 2007, JJF 1190 - 2008

Sampling concentration

≯ 35000 pieces/L

Sampling delay

1~999999 seconds self set

Counting Mode

cumulative value

power supply

High performance specialized lithium battery/AC220V±10%

judgment criteria

ISO14644-1, Static, dynamic, FS-209E

UCL Points

2~50 self setting

sampling location

Self setting can be set with numbers, letters, or a combination of letters and numbers

Particle size distribution error

superior toJJF1190-2008 standard requires ± 30%

Overall dimensions

285×245×285mm

Parameters and models

SENG-3680

SENG-3680B

SENG-3680D

SENG-3680T

SENG-3680Y

automatic alarm

Grade judgment

Grade 100-Million level

Dual use of AC and DC

Storage function

audit trail

Permission Management

Temperature and Humidity Detection

communication interface

RJ45、RS485

RJ45、RS485

sampling flow

2.83L/min

28.3L/min

Screen display mode

5-inch color touch screen

7 Inch color touch screen

battery capacity

built-in 4000mAh

35000mAh

average power consumption

7.5W

90W

Overall dimensions

220*230*150(W*L*H)mm

300*300*300mm

weight

2.8Kg

8.5Kg