YXQ.M-800 pulsating vacuum pressure steam sterilizer
Technical Specifications:
Sterilization room size (mm): 1580 × 650 × 800
Dimensions (mm): 1950 × 1430 × 1950
Effective volume of sterilization room: 0.8m ³
Design pressure: 0.25Mpa
Design temperature: 150 ℃
Work pressure: 0.22Mpa
Working temperature: 134 ℃
Temperature regulation range: 115-135 ℃
Temperature control fluctuation: ± 1 ℃
Sterilization time setting range (seconds): adjustable from 0-9999
Drying time setting range (seconds): adjustable from 0-9999
Thermal uniformity: ± 1 ℃
Pulse frequency: adjustable from 0-99
Control power input power: 0.5KW
Power supply input power: 54KW
Net weight: 1800kg
Note: The working pressure and temperature of this equipment can be adjusted according to the actual working conditions of the user, but should be less than the design pressure and temperature. The appearance should be based on the actual product.
Key Features:
1. The LCD touch screen human-machine interface displays the sterilization information in real-time and dynamically in color throughout the entire operation process, making it more intuitive and clear for users to observe the equipment's operating status. It is also equipped with a printing function, which can record the entire sterilization process.
2. Real time display of temperature and pressure curves, simulation of pressure gauge trends, and real-time monitoring of pressure and temperature changes.
3. The door panel material adopts SUS304 reinforced ribs with Q235B on the outside, and is completely coated with insulation material to minimize heat radiation. The sterilization room switch is driven by a motor to drive transmission mechanisms such as screws and nuts; Using compressed air to tightly press the imported medical silicone sealing strip onto the door panel not only ensures reliable sealing but also greatly reduces the labor intensity of operators compared to manual doors.
4. The door is equipped with a safety interlock device. When there is pressure in the sterilization room, the door automatically locks and cannot be opened to ensure personnel safety. And there is a door control switch, the power can only be turned on when the door is closed in place.
5. Adopt imported high-quality pneumatic valves, solenoid valves, and water ring vacuum pumps. Single chip microcontroller controls electrical components, making them more reliable, stable, and have a longer lifespan.
6. Equipped with fabric sterilization, instrument sterilization, liquid sterilization BD&Helix、 Multiple types of programs such as self sterilization and manual programming.