Welcome Customer !

Membership

Help

Jiangsu Yongsheng Automation Instrument Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products
Product Categories

Jiangsu Yongsheng Automation Instrument Co., Ltd

  • E-mail

    527400968@qq.com

  • Phone

    13365164566

  • Address

    No. 8 Gong'er Road, Jinhu Economic Development Zone, Jiangsu (Second Floor of Comprehensive Building)

Contact Now

YS-100 glass tube level gauge

NegotiableUpdate on 12/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

YS-100 glass tube level gauge is suitable for liquid level display on industrial equipment containers such as chemical and petroleum industries

Product Details

YS-100 glass tube level gauge
 

purpose

This level gauge is suitable for liquid level display on industrial equipment containers such as chemical and petroleum industries.

product structure

The upper and lower valves of the instrument are equipped with flange joints, which are connected to the container to form a connector. The liquid level inside the container is directly indicated by the glass tube.
The structural form is普通型”-D“Thermal insulation type”-WThere are two types, the insulation type is equipped with heating or cooling devices. The device can be heated with steam, and the steam pressure is not greater than0.6MPaWhen cooling, circulating water can be used to adjust the flow performance of the medium. The size of the insulation sleeve connecting pipe isG1/4Internal thread.
Steel balls are installed inside the upper and lower valves. When the glass is damaged due to an accident, the steel balls automatically seal under the pressure inside the container, preventing the medium inside the container from continuing to flow out.
Install limit plug screws on the needle valve at the upper end of the instrument for air release, and also on the needle valve at the lower end for air release, sampling, drainage, and cleaning

Main technical data

1Work pressure: ≤1.0MPa
2
Working temperature: ≤200℃
3
Automatic closing pressure of needle valve: ≥0.2MPa
4
The temperature difference of the heat-resistant rapid change of the glass tube is:180℃
5
The specifications of the instrument are based on the installation center distanceLEstablished:
L
50060080010001100120014001700The2000mm
(Note: Our company accepts orders for special specifications)

Overall Dimension Drawing

structural diagram