Place of Origin

Overview

The YNYSZ-H oleic acid value tester adopts the principle of neutralization method, and automatically completes liquid addition, stirring, titration, and endpoint determination under microcomputer control at room temperature. Adopting a touch sensitive LCD screen to select all working parameters, the human-machine dialogue interface is simple and friendly, easy to operate, and the average oil sample analysis cycle is about 2 minutes. It can display up to 35 sets of measurement results separately and print out multiple parameters related to the measurement.

Product Details

YNYSZ-H Oleic Acid Value Tester

YNYSZ-H Oleic Acid Value TesterIt is a high-precision, fully digital industrial instrument developed by our company's scientific research and technical personnel, based on relevant national standards and utilizing their own advantages through long-term unremitting efforts. The machine is easy to operate and has a beautiful and elegant appearance. The instrument adopts fully automatic digital microcomputer control, with acid-base neutralization reaction as the basic framework for chemical analysis, and is equipped with a self-developed background blank deduction and titration endpoint logic judgment program. Therefore, the precision and accuracy of the measurement are relatively ideal.

The detection of oleic acid value in transformers and steam turbines is a routine inspection item for oil filled electrical equipment in power systems and industrial and mining enterprises, and the acid value tester is a device developed by our company for the determination of the acid value of the above-mentioned oil products. This instrument is based on a single-chip microcomputer and integrates multidisciplinary technologies such as optics, mechanics, electronics, and chemistry. Replacing manual operation not only improves work efficiency, but also reduces the labor intensity of operators and minimizes the harm of organic solvents and other chemicals to the human body.

The oleic acid value automatic tester adopts the principle of neutralization method. At room temperature, the microcomputer controls the automatic completion of liquid addition, stirring, titration, and endpoint judgment. Adopting a touch sensitive LCD screen to select all working parameters, the human-machine dialogue interface is simple and friendly, easy to operate, and the average oil sample analysis cycle is about 2 minutes. It can display up to 35 sets of measurement results separately and print out multiple parameters related to the measurement.

Special reagent bottles are used to contain extraction and neutralization solutions. The extraction solution is stored in a glass container to ensure no plastic additive pollutants. The neutralization solution is stored in polyethylene plastic bottles to avoid the possibility of reaction between caustic potassium and glass. In addition, we have independently developed a neutralization liquid equilibrium gas-solid purification device, which avoids the influence of carbon dioxide and water vapor on the concentration of the neutralization liquid, and avoids accidents where high concentration caustic potassium or caustic soda purification solutions are mistakenly used as titration standard solutions. During the work process, users do not need to touch organic solvents with their hands, reducing the harm of chemicals to the human body.

According to user requirements, the acid value tester can measure 1-3 samples separately.

1. The instrument adopts fully automatic digital microcomputer control, with acid-base neutralization reaction as the basic framework for chemical analysis, and is equipped with a self-developed background blank deduction and titration endpoint logic judgment program. Therefore, the precision and accuracy of the measurement are relatively ideal

2. The instrument is centered around a single-chip microcomputer and integrates multidisciplinary technologies such as optics, mechanics, electronics, and chemistry. Replacing manual operation not only improves work efficiency, but also reduces the labor intensity of operators and minimizes the harm of organic solvents and other chemicals to human health

3. The human-machine dialogue interface is simple and friendly, easy to operate, and the average oil sample analysis cycle is about 2 minutes

1. Linear range: 0.001~0.900 mgKOH/g;

2. Correlation coefficient: 0.999~0.9998

3. Recovery rate: 99~102%

4. MIN resolution: 0.001 mgKOH/g

5. Repeatability: 2-5% (RSD)

6. Measurement accuracy:

The acid value is ± 0.003 mgKOH/g between 0.001 and 0.100 mgKOH/g;

The acid value between 0.100~0.900 mgKOH/g is 2-5% of the indicated value;

7. Display mode: 5.4 "large touchscreen color LCD screen (full Chinese character display)

8. Dimensions: (width) 420 mm × (height) 190 mm × (depth) 340 mm