- Phone
-
Address
No. 27 Binhe Street, Daxing District, Beijing
Beijing Zhongke Environmental Testing Instrument Co., Ltd
No. 27 Binhe Street, Daxing District, Beijing
Xenon lamp weathering aging test chamber
Product UsageThe use of xenon arc lamps that can simulate the full spectrum of sunlight to reproduce destructive light waves in different environments can provide corresponding environmental simulation and accelerated testing basis for scientific research, product development, and quality control. Conduct aging tests by exposing material samples to xenon arc lamp illumination and thermal radiation To evaluate the light and weather resistance of certain materials under high-temperature light sources Mainly used in automobiles, coatings, rubber, plastics, pigments, adhesives, fabrics, etc

1、 Product specifications and technical parameters
| Model Selection | Xenon arc lamp aging test chamber | ||
| SN-100 | SN-150 | SN-500 | |
| Internal dimensions D × W × H | 400×500×500 | 500×600×500 | 500×760×500 |
| Dimensions D × W × H | 850×900×1650 | 750×950×1650 | 1050×1100×1650 |
| temperature range | RT+10℃~100℃ | ||
| humidity range | 50%~98% R.H | ||
| blackboard temperature | 63℃~100℃±3℃ | ||
| temperature fluctuation | ≤±0.5℃ | ||
| Temperature Uniformity | ≤±2℃ | ||
| Humidity fluctuation | +2、-3% R.H | ||
| Xenon lamp source | Air cooled lamp tube (domestic, imported optional) | ||
| Lamp power | 1.8KW/2.5KW/3.0KW (optional) | ||
| Total number of lamps | 1 unit | 2 pieces | 3 pieces |
| Glass filter | 1 piece | 2 pieces | 3 pieces |
| Sample tray size mm | 380×450 | 480×550 | 480×710 |
| rainfall duration | Adjustable continuous rainfall from 0 to 9999 minutes | ||
| Rainfall cycle | Adjustable rainfall interval (intermittent) from 0 to 240 minutes | ||
| Spray cycle | 18min/102min or 12min/48min (spray time/no spray time) | ||
| Rainwater pressure | 0.12~0.15Mpa | ||
| Spray nozzle aperture | Ф0.8mm | ||
| heating power | 2KW | ||
| Humidification power | 1.5KW | ||
| Sample tray size | 450×720mm | ||
| Distance between sample rack and lamp | 230~280mm | ||
| Spectral wavelength | 290nm~800nm | ||
| Continuous adjustable light cycle time | 0~999H、M、S | ||
| Irradiance (optional) |
The average irradiance between 300-400nm is 60W/m2, or 0.51 W/m2 at 340nm (Method 1) The average irradiance between 300-400nm is 50W/m2, or at 420nm it is 1.1 W/m2 (Method 2) The average irradiance between 300-400nm is 60W/m2 to 180W/m2, or 0.51 W/m2 to 1.5 W/m2 at 340nm (Method 1) The average irradiance between (300-400nm) ranges from 50W/m2 to 162W/m2, or from 1.1 W/m2 to 3.6 W/m2 at 340nm (Method 2) |
||
| Power requirements | AC380 (± 10%) V/50HZ three-phase five wire system | ||
SN-100-2022 New Xenon Lamp Weathering Aging Test Chamber
2、 Box material
1. The outer shell is made of (t=1.2mm) A3 steel plate CNC machine tool for processing and forming, and the surface of the shell is sprayed with plastic for a smoother and more beautiful appearance;
2. The inner liner is made of imported (SUS) 304 stainless steel plate;
3. The insulation material adopts high-density glass fiber cotton and high-pressure polyurethane foam insulation;
3、 Heating system:
1. Adopting far-infrared stainless steel high-speed heating electric heater;
2. Independent system for high temperature, humidity, and lighting (without interference with each other);
3. The output power of temperature and humidity control is calculated by a microcomputer to achieve high-precision and high-efficiency electricity efficiency;
4、 Humidification system:
1. The built-in boiler steam humidifier has energy-saving and consumption reducing functions, which can save 70% of energy consumption;
2. Equipped with automatic water level compensation and water shortage alarm system;
3. The humidity control adopts P.I.D+S.S.R, and the system coordinates control with the same channel;
5、 Circuit control system:
1. Control instruments are adopted“7 inchesColor touch screen programmable controller, large screen, easy operation, easy program editing, with R232 communication port, setting display box temperature, box humidity, blackboard temperature and irradiance;

2. Accuracy: 0.01 ℃ (display range);
3. Humidity accuracy: 0.1% (display range);
4. Temperature sensor: PT100 platinum resistance thermometer;
5. Control mode: thermal balance temperature and humidity control mode;
6. Temperature and humidity control is achieved through coordinated control of the same channel using the P.I.D+S.S.R system;
7. It has the function of automatic calculation, which can immediately correct the temperature and humidity change conditions, making temperature and humidity control more accurate and stable;
8. The controller operation interface is available in both Chinese and English, and the real-time operation curve chart can be displayed on the screen;
9. It has a capacity of 100 programs, 100 segments per group, and 999 steps per segment, with a maximum time setting of 99 hours and 59 minutes per segment;
After inputting the data and experimental conditions, the controller has a screen lock function to avoid manual touch and shutdown;
11. Equipped with RS-232 or RS-485 communication interface, it can be programmed on a computer to monitor the testing process and perform functions such as automatic power on/off, curve printing, and data processing;
12. The controller has an automatic screen saver function, which better protects the LCD screen (prolonging its lifespan) during long-term operation;
13. Accurate and stable temperature control, no drift during long-term operation;
14.1s~99h, m, S can set the spray stop time arbitrarily;
15. The instrument displays four screens: box temperature, box humidity, light intensity, and blackboard temperature;
16. Equipped with UVA340 or full spectrum mounted radiometer, real-time detection of irradiance;
17. The independent control time and alternating cycle control program and time for lighting, condensation, and spraying can be set arbitrarily;
In operation or setting, if an error occurs, a warning signal will be provided; Chint's electrical components;
2. Accuracy: 0.1 ℃ (display range);
3. Resolution: ± 0.1 ℃;
4. Temperature sensor: PT100 platinum resistance thermometer;
5. Control mode: thermal balance temperature and humidity control mode;
6. Temperature and humidity control is achieved through coordinated control of the same channel using the P.I.D+S.S.R system;
7. It has the function of automatic calculation, which can immediately correct the temperature and humidity change conditions, making temperature and humidity control more accurate and stable;
8. The controller operation interface is available in both Chinese and English, and the real-time operation curve chart can be displayed on the screen;
reference standard
GB/T16422.2-1999 Plastic laboratory light source exposure test method (Part 2 Xenon arc lamp)
GB/T16422.2-2014 Plastics Laboratory Light Source Exposure Test Method (Part 2 Xenon Arc Lamp)
GB/T8430-98 Textiles - Color fastness test - Xenon arc lamp - Color fastness to artificial light
GB/T1865-2009 Paints, Varnishes, Artificial Air Aging, and Artificial Radiation Exposure
GB/T5137.3-96 Test Method for Radiation Resistance, High Temperature, Moisture Resistance, and Color Aging of Automotive Safety Glass
GB/T2423.24-95 Environmental testing for electrical and electronic products Part 2: Test methods, Test SQ: Simulating solar radiation on the ground
GB/T18244-2000 Aging test method for building waterproofing materials
GB/T3511-2008/ISO 4665:1998
GB/T15102-2006 impregnated film paper veneer artificial board
GB/T12831-91 Vulcanized rubber - Artificial climate (xenon lamp) aging test method
GB/T14522-93 Accelerated Test Method for Artificial Climate of Plastic, Coating, and Rubber Materials Used in Mechanical Industry Products