Product Name: XT 15i Fully Automatic Fat Analyzer
型号: ANKOMXT 15i
Brand: ANKOM
Origin: United States
| Instrument Introduction |
This fully automatic fat analyzer follows Soxhlet extraction, but compared to traditional extraction methods, it operates at higher temperatures, thereby improving extraction efficiency and solvent recovery rate. At the same time, the alloy steel body ensures the safety of operators.
| application |
Suitable for industries such as food, feed, dairy, etc
| Specifications |
Intelligence level: Automatic operation of the entire sample extraction process is completed through programmable computers and controllers to control the driving device
Contains fully automated fat extractionFully automatic solvent recovery and recycling (compared to semi-automatic)
Measurement range%: full range
Range of sample size for one time: 1-3g
Batch processing capacity: 15 pieces
Daily processing capacity: 200 (calculated based on 8 hours of working hours per day)
Batch processing time: 30 minutes (including heating process)
Filter bag structure: three-dimensional structure
Pore size of filter bag: 3um
Temperature range: Room temperature to 90 ℃
Temperature control accuracy: ± 0.1 ℃
Standard deviation of results: 1%
Repeatability accuracy (linear coefficient): 1%
Solvent dosage: 20ml/sample
Recycling situation: 97% of organic solvents can be automatically recycled and used
Safety performance: Low temperature extraction, heating temperature up to 90 ℃, intelligent automatic water circulation cooling, with over temperature shutdown function, automatic shutdown temperature: 105 ℃, solvent recovery process is carried out in a closed state, alloy cylinder body, closed environment, manual operation
Environmental protection situation: Activated carbon ultra trace tail gas adsorption, no need to bring a ventilated kitchen, ordinary standard laboratory is sufficient
Main materials used for the instrument: comply with EX explosion-proof standards and CE electrical standards
| Instrument Features |
Large capacity fat extraction
Up to 15 sample analyses can be completed within 20 minutes
Microcomputer fully automatic control
Solvent recovery and automatic recycling (97%)
Very small footprint