Water cooling system (chiller) for medical accelerators
One:Water cooling system (chiller) for medical acceleratorsThe function of the water cooling system is to cool the heating components of medical accelerators, such as modulators, magnetrons, magnetron electromagnets, circulators, water loads on circulators, and acceleration tubes (including treatment targets), ensuring the normal operation of these heating components.
Secondly:Water cooling system (chiller) for medical acceleratorsThe water cooling system mainly consists of a secondary water heat exchange system, a primary water cooling system, and water pipes and pipelines.
Secondary water heat exchange system: mainly composed of water pump, water tank, water purification equipment, heat exchanger, etc. During system operation, after the secondary water in the water tank is cooled, different heating devices such as medical accelerator tubes, magnetrons, magnetron electromagnets, modulators, circulators, and water loads of circulators are cooled separately through water dividers; The return water flows through the collector and into the water tank to complete the internal circulation of cooling water.
Primary water cooling system: The function of the primary water cooling system is to cool the secondary water, and it is composed of a split structure (integral structure).
3: PID diagram of water cooling system process:

Four:Water cooling system (chiller) for medical acceleratorsSchematic diagram of internal connection with accelerator:

5: Environmental Overview
1)Power Supply:AC voltage 380V ±10%, frequency: 50 Hz ± 10%
2)ambient temperature:The refrigeration unit in a water cooling system can be used outdoors and requires waterproof, dustproof, and antifreeze treatment, making it suitable for operation in-40 ℃~50 ℃, humidity ≤ 95%, atmospheric pressure 5 × 104Pa~11 × 104Pa, altitude ≤ 2000m, protection level not lower than IP54.
3)Environmental humidity: not greater than95% R.H (anhydrous condensation)
4)There is no strong vibration in the environment, and there are no corrosive gases or flammable and explosive gases present
![]()