The walk-in constant temperature environment testing room is a large experimental volume testing equipment, while the walk-in constant temperature environment testing box is a relatively small volume laboratory. However, they are both equipment provided for industrial simulation environment temperature testing of large parts or components. This type of equipment is the strength of Guangdong Aipei Testing Equipment Co., Ltd. It is often produced and serviced. The size, volume, temperature range, and humidity range can be customized according to customer needs, and the temperature fluctuation uniformity and error meet national standards.
Product Details

Walk in constant temperature environment testing roomIt is a testing equipment for large experimental volumes,The walk-in constant temperature environment testing chamber is a relatively small laboratory with a smaller volume,However, they are all equipment provided for industrial simulation environment temperature testing of large spare parts or components.This type of equipment is a strong point produced by Guangdong Aipei Testing Equipment Co., Ltd,Frequent production and service,size,Volume size,temperature range,The humidity range can be customized according to customer needs,The temperature fluctuation, uniformity, and error all meet national standards.

Walk in constant temperature environment testing roomProtection system:

1Overall equipment overheating;

2Fan overheating protection

3The overall equipment is out of phase/Reverse phase;

4Overload of refrigeration system;

5Refrigeration unit overpressure;

6Water shortage protection for humidification system;

7Overall equipment timing;

8Other protections include leakage, operation indication, automatic shutdown after fault alarm, etc.

Walk in constant temperature environment testing roomUsage conditions:

1Installation site

The ground is flat and well ventilated

There is no strong vibration around the equipment

There is no strong electromagnetic field around the equipment

There are no flammable, explosive, corrosive substances or dust around the equipment

Leave appropriate space for use and maintenance around the equipment,

2Power supply conditions

Power requirements:AC380V±10% 50±0.5Hzthree-phase five-wire system

Pre installed power: total power+2.0KW

Require users to configure corresponding capacity air or power switches for the equipment at the installation site, and this switch must be independently used for the equipment (recommended power switch capacity:32A

3Environmental conditions

Ambient temperature5℃~+30℃(24Average temperature within the hour ≤30℃)

ambient humidity:85%RH

4Water supply conditions (wet and hot type and equipment requiring water)

Using purified water, distilled water, and deionized water. resistivity500Ω.m

5Other precautions

Walk in constant temperature environment testing roomOpening the door of the test chamber during the experiment can cause fluctuations in temperature and humidity inside the chamber; If the door is opened multiple times or for a long time during the experiment, or if the test sample emits moisture, it may cause the refrigeration system heat exchanger to freeze and fail to function properly.