Welcome Customer !

Membership

Help

Anhui Kangrong Electric Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Anhui Kangrong Electric Co., Ltd

  • E-mail

  • Phone

  • Address

    No. 20 Renhe Road, Tianchang City, Anhui Province

Contact Now

WSS universal axial bimetallic thermometer

NegotiableUpdate on 05/10
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The WSS universal axial bimetallic thermometer is composed of bimetallic sheets wound in a circular curved shape. When one end expands due to heat, it drives the pointer to rotate, and the working instrument displays the temperature value corresponding to the thermoelectric potential.

Product Details

WSS universal axial bimetallic thermometerApplication:
A bimetallic thermometer is an on-site detection instrument used to measure low to medium temperatures. It can directly measure the temperature of liquid vapor and gas media within the range of -80 ℃ to+500 ℃ in various production processes.

WSS universal axial bimetallic thermometerFeatures:
On site temperature display, intuitive and convenient
Safe and reliable, with a long service life;
The core pulling thermometer can be maintained or replaced for a short period of time without stopping the machine.
There are complete varieties of axial, radial, 135 º, universal and other types, which are suitable for various on-site installation needs.

Working Principle:
A bimetallic thermometer is composed of bimetallic sheets wound in a circular curved shape. When one end expands due to heat, it drives the pointer to rotate, and the working instrument displays the temperature value corresponding to the thermoelectric potential.

Product execution standards:
JB/T8803-1998
GB3836-83
Scale dial nominal diameter: 60, 100, 150
Accuracy level: 1.0, 1.5
Thermal response time: ≤ 40s
Protection level: IP55
The angle adjustment error should not exceed 1.0% of its range
Return difference: The return difference of the thermometer should not exceed the value of the basic error limit
Repeatability: The repeatability limit range of the thermometer should not exceed 1/2 of the basic error limit