WL15A Full Spectrum Spark Direct Reading Spectrometer

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The WL15A full spectrum spark direct reading spectrometer can be used for element quantitative analysis and grade identification of solid metal samples. The optical system adopts a CCD detector, and the spectral range covers typical black and non-ferrous metal materials. The instrument is equipped with a new digital light source and an argon gas flushing spark table. The open spark table is suitable for analyzing samples of different shapes and sizes.

Product Details

Overview
WL15A Full Spectrum Spark Direct Reading SpectroscopyceremonyCan be applied to element quantitative analysis and grade identification of solid metal samples. The optical system adopts a CCD detector with a spectral range covering typical materials of black and non-ferrous metals. It is equipped with a brand new digital light source and an argon gas flushing spark table. The open spark table is suitable for analyzing samples of different shapes and sizes.


technical parameters
project indicator
Detecting the substrate Fe、 Al、 Cu、 Zn、 Ni、 Mg、 Ti、 Co、 Sn、 Pb、Ag Waiting for multiple substrates
Detection Time Depending on the type of sample, generally 10-30s
detector 16 channel high-precision CCD
optical system Pascal Runge structure, Roland light source
wavelength range 140~700nm
focal length 400mmm
pneumatic system Dual gas path and four nozzle gas path system, with built-in flow meter
Light source system Full frequency digital dual pulse excitation light source with variable frequency (200-800HZ) excitation technology
Operating Humidity 20%~80%
Read system High speed 16 bit ADC analog-to-digital conversion, FPGA digital signal controller, industrial USB data acquisition
Inspire the platform The overall material is made of wear-resistant and corrosion-resistant Cr stainless steel, and an intelligent argon gas pressure protection device is activated
Excitation table aperture 13mm
Purity requirements for argon gas 99.999%
Argon inlet pressure 0.5MPa
argon flow rate Excitation flow rate of approximately 3.5L/min
excitation power 600W
Average standby power 100W
size 810(L)×660(D)×380(H)
weight About 80 kg
working power supply (220 ± 20) V AC, (50 ± 1) Hz, protective grounded single-phase power supply
WL15A Full Spectrum Spark Direct Reading SpectrometerFunctional Features
1Optical system
1. Pascal Runge optical structure, vacuum Roland circular grating spectral chamber;
2. Semi submersible vacuum chamber incident lens cleaning device;
3. Micro displacement device for incident slit, high-precision reverse coating incident slit, 10 microns;
4. Grating: Roland circular grating; Engraving line: 2400 lines/mm; focal length: 400mm
5. Wavelength range of the spectral chamber: 120-700nm;
6. Light chamber: ultra-thin constant temperature integrated light chamber, automatic constant temperature of 35 ℃, accuracy ± 0.1 ℃;

Ensure that the instrument can continue to operate normally in any harsh environment, from the laboratory to the furnace site, and even more severe environments.

2Electronic systems
  1. Peak voltage sampling circuit

2. Vacuum measurement circuit of spectrometer optical chamber
3. Adopting ARM chip high-speed USB multi-channel synchronous data acquisition, 16 * 100k synchronous sampling;
4. 16 channel high-precision CCD, Toshiba TCD1304, 3648 pixels, total pixels: 58368 high-resolution full spectrum acquisition
5. AD acquisition: 16 bit single pixel analog-to-digital conversion;
6. Refrigeration device: CCD adopts cooling noise reduction, low noise, and high sensitivity;
7. Signal transmission: Fiber optic digital encoding transmission, strong anti-interference;

3Excitation light source
1. Full frequency digital excitation light source: excitation frequency of 200-800Hz;
2. Digitalization stimulates energy regulation; Dual spark combined discharge: more stable sample excitation;
3. Argon pressure electronic induction protection device: safer excitation, maintaining a constant excitation pressure;
4. Fiber optic command transmission: discharge inductance, capacitance parameter setting;
4Gas path
1. Flow meter pressure regulation: input pressure, 0.25MPa;
2. Flow meter indication adjustment:<5L/min:
3. Set valve with argon constant flow: 0.2L-0.5L/min;
5Sample stand
1. Open sample stage: can adapt to larger sample analysis:
2. Stainless steel excitation platform: durable and long-lasting;
3. Dual gas path excitation device: The sample excitation adopts horizontal and vertical dual gas paths, which is conducive to stable excitation of the sample;
6Vacuum
1. Ultra thin vacuum body: can be equipped with a small vacuum pump;
2. Vacuum degree: Vacuum control is less than 2pa;
3. Intermittent vacuum control device: pumping for 3-5 minutes, can be maintained for>3 hours, ensuring that the light chamber is in a highly sensitive state all year round and avoiding oil and gas pollution;
4. Vacuum pump: equipped with anti backflow device to ensure that the light chamber is not damaged by backflow;
7Power supply
1. Single phase 220V 50Hz, standby power: about 100w; excitation working power: about 600w;
2. Independent power supply for vacuum pump: power: approximately 500w.
8Analysis software

  1. Run all Chinese analysis software on the Windows system, with the ability to switch between multiple language versions according to requirements, set analysis parameters, and activate

Flushing and pre ignition time selection, automatically completed.
2. Automatic data collection, using the least squares principle to automatically fit the working curve.
3. Automatic two-point or single point standardization of analysis curves.
4. Automatic sample control calibration, one analysis curve, can choose multi-point sample control calibration.
5. Output methods available: intensity, intensity ratio, concentration, correction concentration.
6. At the same point, one or more excitations can be repeated, and the results of one or more excitations can be deleted.
7. Automatically calculate the mean, SD, and RSD.
8. Automatically correct interference between elements and background correction.
9. A public standard substance library.
10. The test results can be printed and saved, and the printing format can be arranged by oneself or generated into Word documents for saving.
11. The test results can be transmitted over long distances through networks and computer standard ports.
12. The software comes with fault diagnosis and self-test programs, automatically generating data files that can remotely diagnose instrument faults.

WL15A Full Spectrum Spark Direct Reading SpectrometerBasic Configuration
WL15A Full Spectrum Spark Direct Reading SpectrometerOptional Accessories