Verify the constant temperature bath

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The purpose and characteristics of this standard constant temperature bath are mainly used by metrology departments for the calibration of first - and second-class standard mercury thermometers, Beckman thermometers, industrial platinum resistance thermometers, and standard copper constantan thermocouples. It can also be used as a high-precision heat and cold source for production and scientific research.

Product Details

Verify the constant temperature bathcontrol system

·A temperature control technology developed and self-made using specialized software, equipped with high standard PT100 and fully imported electronic components.

·The software system facilitates the correction of errors between displayed temperature and actual temperature, with a temperature correction resolution of up to 0.001 ℃, ensuring accurate and error free display of temperature values.

·It has over temperature protection, over temperature alarm, and can set over temperature alarm temperature. When over temperature occurs, the load can be automatically cut off.

·The intelligent software has strong temperature stability, and the PID can automatically adjust the optimal parameters according to different media. Special users can also manually readjust the parameters.

·Use software to digitally lock the set values of the control system to prevent unauthorized personnel from operating and ensure the accuracy of experimental data.

·Accurate temperature control ensures rapid and stable temperature inside the working chamber.

·Press the 'Power' button to turn off all functions of the instrument.

Verify the constant temperature bathrefrigeration system

Efficient fully enclosed compressor refrigeration, fast cooling speed, with multiple protections against overheating and overcurrent.

model

temperature range(℃)

temperature fluctuation(℃)

Temperature Uniformity(℃)

digital resolution(℃)

Working slot volume(L)

Work slot opening(mm)

Px -0200M-II

0.~200

0.002~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300

Px -0300M-II

0~300

0.003~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300

Px -10200M-II

-10~200

0.005~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300

Px -10300M-II

-10~300

0.005~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300

Px -20200M-II

-20~200

0.005~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300

Px -20300M-II

-20~300

0.005~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300

Px -30200M-II

-30~200

0.005~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300

Px -30300M-II

-30~300

0.005~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300

Px -40200M-II

-40~200

0.005~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300

Px -40300M-II

-40~300

0.005~0.01/30min

0.005~0.01

0.001

6.5

Φ100x300