Overview
namePlant Living Transformation System
model:Wonbio-PTS
Application fields:
working conditions
1.Power supply voltageDC24V/AC220V±10% 50Hz
2.Ambient temperature0~40℃;
3.relative humidity5-95%;
4.Stay away from conductive dust, explosive, corrosive gases, and sources of strong electromagnetic field interference.
disinfect
After using hazardous substances, please disinfect the equipment using disposable gloves and a disposable cloth75%Clean the instrument and the outer surface of the cavity with alcohol (damp cloth is sufficient, not dripping liquid).
troubleshooting
fault |
reason |
solution |
The display screen is not lit. |
The power switch of the instrument is not turned on |
Turn on the power switch of the instrument |
Poor contact between the instrument and power cord |
Tighten the power cord |
|
Power failure |
Replace the power socket |
|
circuit fault |
Contact the manufacturer's customer service |
|
No response during startup |
Screen button lock |
Unlock the button in the bottom right corner of the screen |
Set the pressure to100 |
Reset pressure |
|
circuit fault |
Contact the manufacturer's customer service |
|
abnormal noise |
The intake valve is not closed |
Rotate the upper and lower intake valves to the closed position |
The inner cabin door is not tightly fastened |
Close the cabin door and rotate it downwards to lock the cabin door |
|
mechanical failure |
Contact the manufacturer's customer service |
Repair Service
1.Non human damage, manufacturer's free one-year warranty; The manufacturer is responsible for repairs outside the warranty period and charges reasonable fees.
2.During the warranty period, users are not allowed to disassemble or assemble parts that are not allowed by the manual on their own, in order to avoid unfamiliarity and the expansion of the fault. Otherwise, the maintenance costs during the warranty period will be borne by the user.
3.If the user wishes to repair it themselves, qualified technicians must be familiar with the structure and circuit principles of the instrument before repairing it. The manufacturer shall not be responsible for the quality of the instrument after repair or the personal safety of the maintenance personnel.
4.Users who need to order parts or repair instruments can contact the manufacturer or authorized agent's after-sales service department.
Transportation and storage conditions
1.Ambient temperature-20℃~55℃;
2.relative humidity5-95%;
3.Handle the instrument gently. If it needs to be moved and transported, it is best to use the original packaging to prevent damage due to severe vibration. During transportation, avoid severe impacts that may cause damage to packaging and products.
User Manual for Plant Living Transformation Systeminstall
Open the specialized packaging of the instrument and carefully check whether the packing list matches the instruments and accessories inside the packaging box. If there are any missing or damaged items, please contact the supplier immediately;
After verifying the accuracy of the inventory, carefully remove the instrument and place it on a flat surface;
Lock the bottom pulley to prevent the instrument from sliding;
Connect the power cord between the instrument and the power socket.
User Manual for Plant Living Transformation SystemInstrument standard configuration list:
Host of Plant Living Transformation System 1Taiwan
4LVacuum pump 1/2a
power cord 1root
Operation Manual 1copy
Warranty Card 1copy
Operating Instructions for Plant Live Transformation System
Instantaneous transformation of plants
Pour the bacterial solution into a 1L glass beaker
Invert the plant and place it on the auxiliary support
Immerse plant leaves into bacterial solution
Place the entire device into the chamber of the plant living transformation system and close the door tightly (rotate the armrest to lock downwards)
Rotate the two air release valves to the "open" position in the lower right corner
Set the pressure on the LCD touch screen to 6-10Ka for 1 minute
Click to start running and hear an alarm sound, indicating that the conversion has been completed
Rotate the release valve to the appropriate position according to the release speed (usually, first rotate to the middle position and then to the bottom)
Plant genetic modification
Cut the explants used for transformation into appropriate shapes and sizes, place them in a culture dish, pour in the appropriate bacterial solution, and cover with a lid
Place the culture dish into the chamber of the plant transformation system and close the door tightly (rotate the armrest to lock downwards)
Rotate the two air release valves to the "open" position in the lower right corner
Set the pressure on the LCD touch screen to 6-10Ka for 1 minute (pressure and time are for reference only; different plants and tissues can explore the conditions and set different values for pressure and time as needed)
Click to start running and hear an alarm sound, indicating that the conversion has been completed
Rotate the release valve to the appropriate position according to the release speed (usually, first rotate to the middle position and then to the bottom)
Precautions
1.Users should carefully read this manual before using this instrument.
2.When using the instrument, ensure that the bottom pulley is in a locked state.
3.This instrument is suitable for short-term and intermittent operation, please do not use it continuously for a long time.
4.Be sure to maintain the "stop" state during operation, check for errors before starting!
5.The instrument must be placed inFlat groundTo prevent the device from sliding during operation.
6.To protect the internal circuits and machinery of the machine, flushing with running water is prohibited, and only a damp cloth can be used for wiping.
7.When the use is over or not used for a long time, the device should be turned off and the power plug should be unplugged.
8.If there is any malfunction during use, the power should be immediately cut off and handed over to professional personnel for handling. Do not disassemble the instrument without confirmation from the manufacturer or authorized agent to avoid secondary damage to the instrument and potential danger to human health.
Maintenance and Care
1.To ensure the continuous stability and accuracy of the instrument, do not place it in an environment with strong magnetic field interference. At the same time, ensure good ventilation and air permeability in the environment.
Attention: Maintenance checks must be completed by the manufacturer, authorized agents, or other qualified personnel;
2.Please use the parts provided by the manufacturer and their authorized agents for the parts that need to be replaced.
clean
Warning: Please turn off the power switch first and unplug the power plug of the instrument.
1.Turn off the power switch and unplug the power plug of the instrument to ensure that it is disconnected from the power supply;
2.Use disposable gloves, dip a disposable cloth in a small amount of water or mild detergent, and clean the instrument and the outer surface of the cavity.