GCMS-QP 2010 Ultra Performance Indicators
1. Ion source
Ionization mode: EI (standard); PCI (option); EI, PCI, and NCI composite source (option)
Ionization voltage: 10 eV to 200 eV
Emission current: 5 uA~250 uA
Filament: Dual filament structure
Ion source temperature: 140 ° C to 300 ° C
Front end open for easy maintenance
2. Mass spectrometry
M/z range: 1.5-1090 amu
Quality analyzer: metal quadrupole with pre quadrupole filter
Sensitivity:
EI full scan, 1pg octafluoronaphthalene m/z 272 signal-to-noise ratio S/N>500:1
PCI scanning sensitivity, 100 pg benzophenone m/z 183 signal-to-noise ratio S/N>500:1
NCI scanning sensitivity, 100 fg octafluoronaphthalene m/z 272 signal-to-noise ratio S/N>500:1
Resolution: 2M unit mass resolution
Quality stability: ± 0.1 u/48 hours
*Scanning speed protocol ASSP: reduces discrimination against high-quality during high-speed scanning and improves sensitivity
*Scanning speed: 20000 u/s
*Small measurement interval/spectrum acquisition speed: 0.01 seconds (* fast 100 Hz)
Color mass spectrometry interface temperature: 50 ° C to 350 ° C
3. Gas Chromatography GC-2010 Plus
Column box operating temperature: 4 ℃~450 ℃
Order of program heating: 20 orders
Injection port temperature: * to 450 ℃
AFC pressure control range: 0~970 kPa
Electronic pressure flow control: Pressure, flow rate, and diversion ratio can be digitally set through the electronic flow control system of *
Injection port: equipped with integrated program heating and cold column head injection functions
4. Detector
Detection unit: Secondary electron multiplier tube with patented deflection lens (to reduce noise) and conversion polarizer
Detector: Patent Active Film Multiplier ®
Dynamic range: 8 x 106
5. Vacuum system
Molecular Turbopump: Differential Dual Pump System: (179 L/sec + 185 L/sec, He)
Pre stage mechanical pump: 30 L/min (60Hz)
Column flow rate: allowable * chromatographic column carrier gas flow rate: 15ml/min(He)
6. Direct injection rod (option)
Temperature range: Room temperature to 500 ° C
Heating rate: 5, 10, 20, 40, 80 ° C/min. (at * output)
Program heating: 5th order program heating
Sample tube size: 1.9 mm inner diameter x 8 mm deep
7. Software (GCMSsolution version 2.6)
Mode: Full Scan, SIM, and FASST (Quickly switch between Scan and SIM measurements to collect data, and obtain spectra from both scan and SIM methods simultaneously)
SIM mode: 64 channels x 128 groups
Energy saving: Ecological Model
Insertion replacement: Easy sTop significantly reduces the maintenance time of the device, allowing for maintenance of the injection port, replacement of glass liners, and isolation pads without stopping the vacuum pump, thereby minimizing instrument downtime
Method:
COAST mode creates an automatic SIM list;
The AART (Automatic Adjustment of Retention Time) function corrects the retention time of compounds based on the retention index and measured values of n-alkanes, without worrying about changes in retention time caused by shortening the chromatographic column or using chromatographic columns with different batch numbers
Spectral retrieval: similarity retrieval, similarity retrieval with retention time index index
Instrument tuning: automatic tuning of all ionization modes (EI, PCI, and NCI)
Mass spectrometry library (option): NIST,Wiley, Pesticide warehouse, essence and spice warehouse, drug warehouse
Database (options): Method Development Kit (Pesticide Residues in Food, EPA 524, EPA 525); Compound composition database software for simultaneous analysis; EPA optional software; VOC analysis software; GC/MS Metabolomics Database
Quality control function: QA/QC function for precise control; System inspection for instrument management; Security features for user management
File format: Compound structure format optimized for GLP
Report format: very flexible report creation and template functionality
8. Installation requirements
Power supply: 50/60 Hz;
Power: Gas phase section 2600 VA (230 VAC); Mass spectrometry section 1000 VA (100~230 VAC)
Environmental temperature: 18~28 ℃
Environmental humidity: 40~70%
9. Volume and weight (excluding computers and printers)
GC/MS:74 kg
Mechanical pump: 10 kg
Used Shimadzu QP2010 gas chromatography-mass spectrometer, used mass spectrometer, imported Shimadzu mass spectrometer
JunqiinstrumentequipmentA new batch of Agilent equipment has arrived, including Agilent 6890 gas chromatograph (equipped with single capillary FID detector; dual capillary FID detector, ECD detector); Agilent 6890N gas chromatograph single capillary FID detector; Equipped with automatic liquid sampler and sampling tower, with a hundred position sample tray; Agilent 7694 headspace sampler (44 stations fully automatic, 6 stations sample heating); Agilent 6890N+5973 gas chromatography-mass spectrometer GCMS Shimadzu QP2010 GCMS, Agilent LC-MS, AB (API) GC-MS, etc.
Before making a purchase, be sure to contact the small shopHelp you find the item you need at a suitable price
contact information 18017027751
www.sh-junqi17.com
Specific configuration price, phone contact
Junqi Instrument's new chromatography equipment rental service is currently limited to customers in Jiangsu, Zhejiang, and Shanghai. The rented equipment includesAgilent 5890P gas chromatograph Agilent 6890 gas chromatograph Agilent 6890N-5973 gas chromatography-mass spectrometer Shimadzu QP-2010 gas chromatography-mass spectrometer details consultation Junqi Instrument 18017027751
Updated on September 2, 2020 at 10:22:27