Welcome Customer !

Membership

Help

Junqi Instrument Equipment (Shanghai) Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products
Product Categories

Junqi Instrument Equipment (Shanghai) Co., Ltd

  • E-mail

  • Phone

    18017027751

  • Address

    No. 293-295 Yexie Yexiao Road, Songjiang District, Shanghai

Contact Now

Pesticide residue rapid detector (96 channel)

NegotiableUpdate on 02/19
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The ZYD-NP series high-throughput pesticide residue detector is designed based on the acetylcholinesterase inhibition rate method in the national standard, and can quickly and quantitatively detect the residual pesticide content in food. Used for detecting organic phosphorus and amino methyl ester pesticide residues in vegetables, fruits, tea, grains, water, etc. Suitable for agricultural testing centers, production bases, farmers' markets, supermarkets, hygiene, environmental protection, hotels and other fields at all levels.
Product Details

Introduction:

* 5 Inch full Chinese color large screen LCD display

* ☆ Touch screen operation, convenient and intuitive

* The light source adopts ultra high brightness light-emitting diodes, which have lowAdvantages such as high power consumption, high reliability, and fast response speed.

* ☆ Adopting a closed-loop light source automatic calibration system,Avoiding prolonged use or changes in external conditionsThe phenomenon of excessive or insufficient light source caused by chemical transformation should be avoidedProve that the light source is always working in the best condition.

* ☆ Light source preheating and constant temperature management system, effectively avoidingDrift free, ensuring stability for long-term measurements.

* ☆ Automatic calibration of instruments0And100%No needManually perform this calibration operation.

* 8Channel optical path measurement system

* Accurate automatic injection positioning system ensures testingThe accuracy of the results

* ☆ Equipped with vibration plate function, the vibration speed and time can be adjustedAdjustment

* Equipped with instrument self checking function

* USBCommunication structure, convenient for connecting to computersData processing

* Internal miniature thermal printer

parameter

Light source wavelength:410nm

* Detection channel:96 Simultaneous channel inspectionTesting;

* Reaction time:1 Minutes or3 divideClock selection;

* Detection limit:0.055.0mg/kgOrganic phosphorus and amino acid estersClass);

* Test scope:03.5A

* Resolution:0.001A

* Accuracy: ±1%02A);

* Linear error: ±0.1%0 2A);

* Repeatability: ±0.005A0 2A);

* Stability: ≤0.005A

* Storage: Large capacity storage, capable ofStorage about300000 Group measurement rawdata

* Interface: Embedded Micro ThermistorPrinter, large color screenLCDChinese display.

Updated on September 2, 2020 at 10:22:42