Welcome Customer !

Membership

Help

Ningbo Yixin Optoelectronics Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Ningbo Yixin Optoelectronics Technology Co., Ltd

  • E-mail

    easson@aieasson.com

  • Phone

    13616887002

  • Address

    No. 10 Xincheng Road, Cicheng Town, Jiangbei District, Ningbo City, Zhejiang Province

Contact Now

UV aging test chamber

NegotiableUpdate on 12/24
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
This experimental equipment simulates the damage caused by solar ultraviolet light, and conducts weather resistance testing on the tested material by exposing it to controlled high temperature light. It uses ultraviolet lamps to simulate the radiation effect of sunlight, and simulates dew and rain through condensation and water spraying. In just a few days or weeks, the ultraviolet irradiation equipment can withstand the damage that takes months or even years to occur outdoors, including fading, color changes, loss of luster, powdering, cracking, wrinkling, bubbling, brittleness, strength reduction, oxidation, etc. Its test results can be used to select new materials, improve existing materials, or evaluate materials
Product Details

Equipment usage:
This experimental equipment simulates the damage caused by solar ultraviolet light, and conducts weather resistance testing on the tested material by exposing it to controlled high temperature light. It uses ultraviolet lamps to simulate the radiation effect of sunlight, and simulates dew and rain through condensation and water spraying. In just a few days or weeks, the ultraviolet irradiation equipment can withstand the damage that takes months or even years to occur outdoors, including fading, color changes, loss of luster, powdering, cracking, wrinkling, foaming, brittleness, strength reduction, oxidation, etc. Its test results can be used to select new materials, improve existing materials, or evaluate changes in material formulations.

Product Introduction

Sample restriction conditions:

lThis experimental equipment prohibits:

lTesting and storage of samples of flammable, explosive, and volatile substances

lTesting and storage of corrosive substance samples

lTesting or storage of biological samples

lTesting and Storage of Strong Electromagnetic Emission Source Samples

Technical Indicators:
1.Volume and size:

inner boxSize(mm):1140×640×690DWidth xWDeep xHHigh)

Outer box size(mm):1300×700×1630DWidth xWDeep xHHigh)

2.The Sun's Eye:

Range of irradiance:0.31.0W/m2adjustable

Extend the lifespan of the lamp tube

3.UV lamp tube parameters:

Lamp model:UVA-340(315400nm)TheUVB-313(280315nm)Optional, this plan requires8Support the lamp tube. Standard configuration isUVA-340

Lamp brand: AmericanATLAS

Center distance of the lamp:70mm

Distance between test rack and lamp tube: Plate type base frame test board (adjustable up and down)

Radiation intensity: output according to the set light intensity

4.Temperature and humidity indicators:

Temperature range: normal temperature+1070℃ adjustable

Temperature fluctuation:±0.5

Temperature resolution:0.1

5.Blackboard thermometer:

Measurement range:080

Tolerance:±3

6.Temperature control method:PIDSelf setting temperature control method

7.Technical features:

lProvide comprehensive and detailed real-time monitoring and curve recording of important experimental parameters such as irradiance, temperature, and time (controller with recorder function), and generate reports and print curves through the computer.

lCan be irradiated separately/condensation/Water spray test, can also be programmed with any combination of irradiation/condensation/Water spray test.

ladoptOYOindustrial-grade7The inch true color touch screen control system has good stability, monitoring, operability, and maintainability.

lProvide lifespan reference for imported ultraviolet lamps with expensive materials (lamps are consumables with a lifespan of approximately)1600Hours), convenient for users to confirm the replacement time of the lamp tube, saving usage costs. Through the irradiance measurement and control device (solar eye), the irradiance can be measured and controlled to automatically stabilize at the set value, greatly extending the service life of the lamp.

lAdopting automated water supply and drainage design, combined with high cost high-power mercury water, to ensure the flow rate and uniformity of water spray.

lMost of the main components are from international brands, while a small portion are from domestic brands, which improves the safety and reliability of the product. (Please refer to the attached list of main accessories for details.)

lThe installation and wiring of the electrical part of the equipment shall be carried out in accordance with international electrical standards.

lImprove professional after-sales service.

8.meet the standard

lGB/T14552-2008National Standards of the People's Republic of China--Plastics, coatings, and rubber materials used in mechanical industry productsThe Accelerated Test Method for Artificial ClimateaFluorescent ultraviolet/condensation test method.

lGB/T16422.3-1997 GB/T16585-96Related analysis methods.

lGB/T16585-1996National Standards of the People's Republic of ChinaTest Method for Artificial Climate Aging of Vulcanized Rubber (Fluorescent UV Lamp).

lGB/T16422.3-1997Design and manufacturing standards for corresponding standard clauses such as "Plastic Laboratory Light Source Exposure Test Method".

lCompliant with international testing standards:ASTM D4329TheISO 4892-3TheISO 11507TheSAE J2020Waiting for the current UV aging test standards.

IIFunctional Area Introduction:

1Radiation system (simulated sunlight):

llight source8The rated power is40WImported ultraviolet fluorescent lamps from the United States ATLAShaveUVA-340andUVB-313The light source is available for users to choose and configure.

lSolar Eye System:

Changes in light intensity can affect the rate of material aging, while differences in spectra may simultaneously affect both the rate and type of material aging

Generate impact. Accurate and reproducible test results can only be obtained by controlling irradiance through the solar eye system.

The solar eye system is a precise closed-loop control system that can automatically maintain the set light intensity through a feedback loop. control

The device can continuously monitor the intensity of ultraviolet light and compensate for changes in light intensity caused by lamp aging or any other factors by adjusting the output of the lamp. The reproducibility and repeatability of the test results of the solar eye irradiance control system are better than those of manual irradiance control, and can effectively extend the life of the lamp tube.

The sun eye is installed in the distribution cabinetUVIn the loop, the solar eye monitors the light intensity of one (two) set of lamps and transmits the data to the controller. The controller compares the measured irradiance value with the set value, and then instructs the power system to adjust the lamp output to maintain the set light intensity value.



a)UVA-340: Can emit from315nmto400nmThe wavelength of the sunlight in between, and its peak radiation is at340nmHere.UVA-340Comparative testing of different experiments is particularly useful, as it can effectively simulate sunlight in the short wave ultraviolet band.

UVA-340Reference point for irradiance setting:0.68W/m2Equivalent to the maximum intensity of sunlight at noon in summer, obtaining quick results.


b)UVB-313:The wavelength range is280nmto315nmbetween. The peak is at313nmHere.UVB-313compareUVA-340The material aging caused by light tubes is faster, but their shorter wavelengths than the cut-off point of sunlight may produce unrealistic results for many materials.

UVB-313Reference point for irradiance setting:0.67W/m2yesUVB-313Typical light intensity of the lamp tube, very fast testing, application Testing of materials with good quality control or durability.


lUVCycle temperature control:

UVThe components of the circulatory system include a blower, an air heater, an air distribution system, and a blackboard temperature sensor.

a)Blower: installed at the air duct of the lower box of the equipment, throughout the entireUVIn the cycle, the blower works continuously.

b)Air heater: Installed in the air duct above the blower, it is used to heat the air blown out by the blower when needed.

c)Air distribution system: The air blown out by the blower enters the laboratory through the air duct at the center of the water tray, and is then guided by air to fill the entire laboratory.


2Condensation system (simulated dew):

QUVThe condensation system is essentially the evaporation of water vapor confined within the experimental chamber by the sample of the experimental chamber. During the condensation cycle, water is heated by the heating element in the water tray, and the test chamber is filled with steam100%When the humidity reaches equilibrium, water vapor continuously condenses on the sample and finally flows back into the water tray.



3Spray system (simulating rainwater):

QUVThe sprinkler system is equipped with a high-power water pump,5A nozzle (on each side)4It consists of a connecting pipe, a control valve, and a drainage part. The nozzle is installed between the lamp tubes, and in the spray cycle, water is sprayed onto the sample to produce thermal shock or erosion effects.


4Structural characteristics of the box:

lThe box is made with full CNC technology, smooth lines, and a beautiful and elegant appearance

lThe material of the whole box isSUS304Advanced stainless steel plate

lStainless steel dense hole typesampleProduct rack

lDoor type opening devices are used at all openings of the test box casing, making it easy to open

lFor ease of movement, casters are installed at the bottom of the instrument and fixed in position before useUPosition of the ring fixed testing machine

5Control panel: (Specific buttons and indicator lights depend on the actual configuration)

lOYO1200 7Inch programmable display screen

lRS232communication port

lOperation button

lAlarm indicator light

lIrradiation timer

6Laboratory:

lUVLamp tube, blackboard temperature sensor.

lHeating and humidification device

lInlet and outlet, sprinkler system

7Mechanical room

lInlet and outlet system

lSpray waterway

lCondensed water circuit

lBlower, air heater

lalarm device

8distribution control cabinet

lDistribution board, main power circuit breakerCooling fanwait

9heater

lHeater material: stainless steel304Fin heat dissipationelectricityheating tube

lHeater control method: contactless equal period pulse width modulation,SSR(Solid state relay)

10humidifier

lHumidification method:shallowBasin heating

lHumidifier material: stainless steel armor

lHumidifier control method: contactless equal period pulse width modulation,SSR(Solid state relay)

lHumidifier device: water level control device, heater anti dry burning device

11Electrical control system:

lController (model):Programmable touch screen controllerOYO1200

lmonitor: 7Inch true color touch screen

lOperation mode: program mode, fixed value mode

lSetting method: Chinese menu, touch screen input

lProgram Capacity: Memory Program Group120Group,

lEach program group contains program segments120Segment, all loops, maximum9999return

lBlackboard temperature monitoring: using standardPt-100Blackboard temperature sensor,Accurately control the surface temperature of the sample during the testing process.

lAuxiliary functions: fault alarm and cause, processing prompt function, power-off protection function, timing function(Automatic start and automatic stop operation)Self diagnostic function.


12 and OYO1200Controller operation (please refer to the controller manual for detailed operation):

lInitial screen

Users can switch to other screens through the home directory screen, as shown in the following figure.


lFixed value operation function.

Users can use this feature to achieve continuous UV testing. In the customized operation screen, users can manually control the opening and closing of various functions. On this screen, temperature setting can be achieved with automatic temperature calculation function.


ChengformulaOperation function.

On the program control screen, users can automatically run the device according to the edited program. When the program is running, the device has functions such as maintaining skip segments. And it can intuitively observe basic information such as the total running time of the program, the running time of program segments, and the program number. At the same time, the operating status of each system will also be displayed intuitively on the screen.


lDocument management function

This device has powerful curve display, document recording, management, and storage functions.

3、 List of main accessories

serial number

Accessory name

quantity

Model

1

controller

1

touch screen,7-inch true color, resolution 800*480

2

Irradiance solar eye

4

UVA340 / UVB313(Industrial grade, with filter)

3

UV lamp

8

standard configurationUVA34040W, Diameter 38mm, Long1200mm)

4

AC contactor

5

CJ*2-18102), CJ*2-09103 of them)

5

overtemperature protection

1

50~320degree

6

solid state relay

2

3-32direct current / 70-280V communication

7

temperature sensor

1

PT100Armor(-100~250Degree)

8

circuit breaker

1

NXB-63 C25400V communication/ 50Hz6000A

9

circulating fan

3

AC220V/50Hz8Leaf steel shaft

10

heater

1

AC220V/50Hz1.5KW

11

humidifier

2

AC220V/50Hz 1.5KW

12

Spray booster water pump

1

AC220V/50Hz370W

4、 Safety protection device:

lFan duct over temperature protection

lShallow water heating tube anti dry burning protection

lWater tank water shortage alarm protection

lWater tank and waterway fault alarm protection

lShort circuit protection, overcurrent protection

5、 Usage conditions (guaranteed by the user for the following)Condition):

1. Venue:

lThe ground is flat, well ventilated, and free of flammable, explosive, corrosive gases, and high concentration dust

lThere is no strong electromagnetic radiation source nearby

lNo direct sunlight or direct radiation from other heat sources

lThere are water supply and drainage devices near the equipment

lGround area and load-bearing capacity of the venue: not less than234 × 353cm and 500kg/m2

lLeave appropriate maintenance space around the equipment

2. Environmental conditions:

ltemperature5 ℃~32 ℃, preferably installed in air-conditioned rooms. This can keep the room temperature between 21 ℃ and 27 ℃.

lrelative humidity≤ 85%RH

lAir pressure:86~106kpa

3. Electricity requirements:

lEquipment power:3.5KW, Voltage: 220V ± 10%, single-phase three wire system, 50/60Hz

7、 Documents and attachments provided:

lProduct factory certificate of conformity1 copy

lProduct User Manual1 book

lProduct Warranty Card1 sheet