- Phone
-
Address
Hongchuang Technology Park, No.1 Baoda Road, Suzhou Industrial Park
Suzhou Guangjun Experimental Instrument Co., Ltd
Hongchuang Technology Park, No.1 Baoda Road, Suzhou Industrial Park
Product Features
Adopting a touch based graphical control interface, the operation is simple.
The impact method applies the air path switching method to import the temperature into the testing area for conducting cold and hot impact tests.
During high or low temperature impact, the maximum time can reach 999H and the maximum cycle period can reach 9999 times.
The system can perform automatic cyclic impact or manual selective impact.
The cooling system adopts a binary freezing system, with fast cooling effect, and the cooling method is air-cooled or water-cooled.
Implement standards and test methods that meet the requirements
GJB150.5 Temperature Shock Test
GJB360.7 Temperature Shock Test
GB/2423.22 Temperature shock test, etc
Selection Table for Two Impact Test Boxes
| type number | GLT2-A | GLT2-B | GLT2-C | GLT2-D | GLT2-E | GLT2-F | GLT2-G | |
| Inner container size cmW * H * D | 30*30*30 | 40*40*40 | 50*40*40 | 60*50*50 | 60*60*60 | 80*90*70 | 100*100*100 | |
| Test Method | Two boxes of movable items | |||||||
| performance parameters | temperature range | Low temperature (-60 ℃, -50 ℃, -40 ℃, -30 ℃, -20 ℃) to high temperature (+85 ℃,+100 ℃,+120 ℃,+150 ℃) | ||||||
| temperature error | Cold storage room regenerator ≤±2℃ | |||||||
| volatility | ≤± 0.5 ℃ (constant temperature state) | |||||||
| impact time | Impact recovery time within 5M Basket transfer time within 10 seconds | |||||||
| control system | Controller | Touchscreen integrated temperature shock controller ( Cold storage heat storage pilot zone Three zone independent display) | ||||||
| precision range | Setting accuracy: temperature ± 0.1 ℃, resolution of analysis: ± 0.01 ℃ | |||||||
| refrigeration system | Mechanical binary compressor unit (cascade refrigeration) | |||||||
| heating system | Nickel alloy high-speed electric heater | |||||||
| circulatory system | Imported motor, scientific air circulation design, no temperature blind spots indoors | |||||||
| Defrosting system | There are two types of defrosting systems: automatic defrosting system and manual defrosting | |||||||
| Materials used | outer box material | SUS304 high-grade stainless steel shell (optional cold-rolled sheet spray treatment) | ||||||
| Inner box material | SUS304 high-grade stainless steel plate (all three boxes are) | |||||||
| sealing material | Imported double-sided high-quality silicone sealing strip | |||||||
| thermal insulation material | High quality ultra-fine glass fiber | |||||||
| equipment power supply | AC 380V 50Hz | |||||||
| safety device | Power overload protection, leakage protection, short circuit protection, compressor protection, grounding protection, over temperature protection, voice alarm, etc | |||||||