Thin film breakdown voltage tester

NegotiableUpdate on 01/12
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
LJC - (150000 volts) is mainly used for measuring the power frequency breakdown voltage test of solid insulation materials, such as insulation paper and cardboard, laminated board, insulation tube, plastic, film, rubber, insulation paint and other single materials, as well as conducting breakdown voltage and withstand voltage tests using liquid or gas as impregnating agents. Thin film breakdown voltage tester
Product Details
high voltageThin film breakdown voltage tester
1Thin film breakdown voltage testerpurpose
LJC - (150000 volts) is mainly used for measuring the power frequency breakdown voltage test of solid insulation materials, such as insulation paper and cardboard, laminated board, insulation tube, plastic, film, rubber, insulation paint and other single materials, as well as conducting breakdown voltage and withstand voltage tests using liquid or gas as impregnating agents.
2、 Voltage breakdown device
I. Basic Parameters
1. Input voltage: AC380V/AC220V
2. Output voltage: AC 0-150 kV DC 0-150 kV
3. Boosting method: continuous boosting, gradual boosting, withstand voltage
4. The test transformer requires independent oil immersion, with a measuring coil and a capacity of 20KVA. It is used for high-voltage measurement with a voltage transformer to directly measure the electrode voltage. The measuring coil end and the measuring end of the transformer should be unified, and the voltmeter can be connected arbitrarily
5. High voltage classification: 0-150kV 0-20kV
6. Breakdown voltage: 0-150kV 0-20kV
7. Boosting rate: 0-150kV: 100V/s~5kV/s, stepless speed regulation can be set arbitrarily
0-20 kV: 20V/s~2kV/s stepless speed regulation can be set arbitrarily
8. Voltage measurement accuracy: (10% --100% FS) ≤ 2%
9. Test method: Under power frequency, test in: 1. Insulation sample in air test 2. Insulation sample immersed in oil (at room temperature or in a high-temperature oil cylinder) test
10. Boosting method: 1. Constant speed boosting 2. Gradient boosting 3. Voltage withstand test
11. Stop detection method: 1. Voltage stop detection 2. Current stop detection
12. Test medium: 1. In air 2. In transformer oil (at room temperature or in high-temperature oil cylinders)
13. Overcurrent protection device: Cut off the power supply within 0.2 seconds when the sample breaks down
14. Leakage current selection: (0-100mA) can be freely set
15. The equipment mainly consists of high-voltage test transformers, contact voltage regulators, servo speed control systems, acquisition systems, control (manual and computer) systems, automatic discharge, and other parts
16. Manual control console, computer control console, booster, transformer, low temperature box, oil cylinder, and discharge ball pole all have a safe distance (connected by wires)
II. Control parameters
1. Voltage breakdown device 20, 150KV transformer capacity and measurement are independent systems
2. Control mode: Both 20KV and 150KV breakdown devices have computer control and manual control consoles (two independent systems that can be switched between each other, among which the measuring voltmeter of the manual console should have the function of retaining the maximum test voltage reading)
3. Computer control
3.1. During the experiment, the experimental curve can be dynamically drawn, and the experimental curve can be overlaid with multiple colors for comparison
3.2. Can edit and modify experimental data, flexible and applicable
3.3. The experimental conditions and test results can be stored as data
3.4. Can manually select the validity of curve data in a set of experiments
3.5. The experimental result data can be imported into EXECL
3.6. The software operation interface allows operators to operate freely and has greater affinity.
3.7. The software should include:
Test method: The selection of "breakdown", "withstand voltage", "gradient withstand voltage" or test method can be carried out
Peak voltage drop: used to determine whether a material has broken down. An instantaneous drop in voltage exceeding this value is considered breakdown
Stop current judgment: used to determine the current at breakdown, exceeding this current is considered breakdown
Initial voltage: used for withstand voltage and gradient withstand voltage tests, where the voltage is raised at the beginning of the test
Step by step voltage: used for gradient withstand voltage test, setting the gradient value of the boost voltage
Step by step time: used for gradient withstand voltage test, set at the corresponding gradient withstand voltage time
4. Manual control (without using a computer)
4.1. The panel has selectable options for "breakdown", "withstand voltage", and "gradient withstand voltage"
4.2. Emergency button and pressure indicator light
4.3. The test voltage is digital display type
4.4. There is a time control table for pressure resistance
4.5 The control mode is manual button to apply voltage or stop
III. Instrument design and application comply with standards
Meets GB1408.1, GB/T1695, GB/T3333, GB 12656, AST D149, AST D 876, IEC60243-1:1998
IV. Safety Protection
1. Overcurrent protection: Low voltage side overcurrent protection, protecting the safe operation of high-voltage transformers, high-voltage side overcurrent protection, protecting electrode surfaces from spark corrosion, exceeding preset leakage current, and cutting off high-voltage output
2. High voltage power-off protection: Automatically cut off high voltage output beyond the range, manually shut down high voltage output during the experiment, and cut off high voltage output if the voltage drop exceeds the preset value
3. Output zero protection: Before the experiment starts, if the high-voltage output is not at zero position, a prompt will be given. If the high-voltage output is not at zero position, it will be forced back to zero
4. Short circuit protection: High voltage output short circuit, automatic cut-off of output, low voltage input short circuit, automatic power-off
5. Software protection: When the high voltage preparation switch is pressed, the high voltage indicator light will light up and the voltage regulator will reset (high voltage output will be zero)
6. Leakage protection: independent grounding protection, leakage protection switch
7. The control panel is isolated from other components by a safety fence and has a chain safety door
8. The control panel has a safety warning bell button and an interlocking device with the boost button
9. The test bench and high-voltage equipment components have shielding mesh isolation
VII. System Composition
1. Intelligent control console (manual control and computer control)
2. 150kV/20kVA oil immersed test transformer
3. 150kV continuously variable speed regulator
4. Testing system
5. Protective device
6. Current and voltage limiters
7. Test electrode
8. Test oil tank
3、 Oil cylinder
Technical Specifications
1. Temperature range: Room temperature to 150 ℃
2. Size: Suitable for testing with a 150kV voltage breakdown tester (with a diameter of not less than 1.2m and a height of not less than 1m)
3. Material: resistant to transformer oil and high temperature
4. Heating method: Electric heating, heating rod coiled around the inner wall of the oil cylinder
5. The oil cylinder is equipped with an insulated cover and a hole reserved in the middle for introducing the high-voltage lead of the breakdown instrument. The outside of the oil cylinder is equipped with grounding and low-voltage leads (with crocodile clips), and the inside of the oil cylinder is equipped with an insulation frame for placing electrodes